EUVIMEDEUVIMED entdecken
Open Access

Simulation study of protoacoustics as a real‐time in‐line dosimetry tool for FLASH proton therapy

Kaitlyn Kim, Prabodh Kumar Pandey, Gilberto Gonzalez, Yong Chen, Liangzhong Xiang

Medical Physics · 2024

DOI: 10.1002/mp.16894 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Kaitlyn Kim und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of California, Irvine, California, USA. Hauptthemen laut MEDLINE: Proton Therapy, Radiometry, Monte Carlo Method, Radiotherapy Dosage. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 56 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Chao Family Comprehensive Cancer Center, American Cancer Society, NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Applying ultra‐high dose rates to radiation therapy, otherwise known as FLASH, has been shown to be just as effective while sparing more normal tissue compared to conventional radiation therapy. However, there is a need for a dosimeter that is able to detect such high instantaneous dose, particularly in vivo. To fulfill this need, protoacoustics is introduced, which is an in vivo range verification method with submillimeter accuracy.

Purpose

The purpose of this work is to demonstrate the feasibility of using protoacoustics as a method of in vivo real‐time monitoring during FLASH proton therapy and investigating the resulting protoacoustic signal when dose per pulse and pulsewidth are varied through multiple simulation studies.

Methods

The dose distribution of a proton pencil beam was calculated through a Monte Carlo toolbox, TOPAS. Next, the k‐Wave toolbox in MATLAB was used for performing protoacoustic simulations, where the initial proton dose deposition was inputted to model acoustic propagations, which were also used for reconstructions. Simulations involving the manipulation of the dose per pulse and pulsewidth were performed, and the temporal and spatial resolution for protoacoustic reconstructions were investigated as well. A 3D reconstruction was performed with a multiple beam spot profile to investigate the spatial resolution as well as determine the feasibility of 3D imaging with protoacoustics.

Results

Our results showed consistent linearity in the increasing dose‐per‐pulse, even up to rates considered for FLASH. The simulations and reconstructions were performed for a range of pulsewidths from 0.1 to 10 μs. The results show the characteristics of the proton beam after convolving the protoacoustic signal with the varying pulsewidths. 3D reconstruction was successfully performed with each beam being distinguishable using an 8 cm × 8 cm planar array. These simulation results show that measurements using protoacoustics has the potential for in vivo dosimetry in FLASH therapy during patient treatments in real time.

Conclusion

Through this simulation study, the use of protoacoustics in FLASH therapy was verified and explored through observations of varying parameters, such as the dose per pulse and pulsewidth. 2D and 3D reconstructions were also completed. This study shows the significance of using protoacoustics and provides necessary information, which can further be explored in clinical settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyRadiometryMonte Carlo MethodRadiotherapy DosageAcousticsTime FactorsComputer SimulationFeasibility StudiesHumans

Stichwörter der Autor:innen

flash proton therapyin vivo dosimetryionoacousticsprotoacoustics

Autor:innen und Institutionen

  • Department of Biomedical Engineering, University of California, Irvine, California, USA.
  • Department of Radiological Sciences, University of California, Irvine, California, USA.
  • Department of Radiation Oncology, University of Oklahoma College of Medicine, Oklahoma City, Oklahoma, USA.
  • Beckman Laser Institute & Medical Clinic, University of California, Irvine, California, USA.

Förderung

  • Chao Family Comprehensive Cancer Center
  • American Cancer Society
  • NCI NIH HHS
  • National Cancer Institute

Zitieren

APA

Kim, K., Pandey, P. K., Gonzalez, G., Chen, Y., & Xiang, L. (2024). Simulation study of protoacoustics as a real‐time in‐line dosimetry tool for FLASH proton therapy. Medical Physics. https://doi.org/10.1002/mp.16894

Vancouver

Kim K, Pandey PK, Gonzalez G, Chen Y, Xiang L. Simulation study of protoacoustics as a real‐time in‐line dosimetry tool for FLASH proton therapy. Medical Physics. 2024. doi:10.1002/mp.16894

BibTeX
@article{Kim2024,
  author = {Kim, Kaitlyn and Pandey, Prabodh Kumar and Gonzalez, Gilberto and Chen, Yong and Xiang, Liangzhong},
  title = {Simulation study of protoacoustics as a real‐time in‐line dosimetry tool for FLASH proton therapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16894}
}
Kostenlos starten →