EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane‐parallel ionization chamber

Benjamin Gebauer, Kilian-Simon Baumann, Hermann Fuchs, Dietmar Georg, Brad M Oborn, Hui-Khee Looe, Armin Lühr

Medical Physics · 2024

DOI: 10.1002/mp.16797 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Benjamin Gebauer und 6 weiteren Autor:innen. Beteiligte Einrichtung: OncoRay National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dre…. Hauptthemen laut MEDLINE: Protons, Proton Therapy. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: European Metrology Programme for Innovation and Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The combination of magnetic resonance imaging and proton therapy offers the potential to improve cancer treatment. The magnetic field (MF)‐dependent change in the dosage of ionization chambers in magnetic resonance imaging‐integrated proton therapy (MRiPT) is considered by the correction factor , which needs to be determined experimentally or computed via Monte Carlo (MC) simulations.

Purpose

In this study, was both measured and simulated with high accuracy for a plane‐parallel ionization chamber at different clinical relevant proton energies and MF strengths. Material and methods The dose‐response of the Advanced Markus chamber (TM34045, PTW, Freiburg, Germany) irradiated with homogeneous 10 10 cm quasi mono‐energetic fields, using 103.3, 128.4, 153.1, 223.1, and 252.7 MeV proton beams was measured in a water phantom placed in the MF of an electromagnet with MF strengths of 0.32, 0.5, and 1 T. The detector was positioned at a depth of 2 g/cm in water, with chamber electrodes parallel to the MF lines and perpendicular to the proton beam incidence direction. The measurements were compared with TOPAS MC simulations utilizing COMSOL‐calculated 0.32, 0.5, and 1 T MF maps of the electromagnet. was calculated for the measurements for all energies and MF strengths based on the equation: , where and were the temperature and air‐pressure corrected detector readings with and without the MF, respectively. MC‐based correction factors were determined as , where and were the doses deposited in the air cavity of the ionization chamber model with and without the MF, respectively. Furthermore, MF effects on the chamber dosimetry are studied using MC simulations, examining the impact on the absorbed dose‐to‐water () and the shift in depth of the Bragg peak.

Results

The detector showed a reduced dose‐response for all measured energies and MF strengths, resulting in experimentally determined values larger than unity. For all energies and MF strengths examined, ranged between 1.0065 and 1.0205. The dependence on the energy and the MF strength was found to be non‐linear with a maximum at 1 T and 252.7 MeV. The MC simulated values agreed with the experimentally determined correction factors within their standard deviations with a maximum difference of 0.6%. The MC calculated impact on was smaller 0.2 %.

Conclusion

For the first time, measurements and simulations were compared for proton dosimetry within MFs using an Advanced Markus chamber. Good agreement of was found between experimentally determined and MC calculated values. The performed benchmarking of the MC code allows for calculating for various ionization chamber models, MF strengths and proton energies to generate the data needed for a proton dosimetry protocol within MFs and is, therefore, a step towards MRiPT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ProtonsProton TherapyRadiometryMonte Carlo MethodWaterMagnetic Fields

Stichwörter der Autor:innen

ionization chamber dosimetrymagnetic fieldsprotons

Substanzen

ProtonsWater

Autor:innen und Institutionen

  • OncoRay National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany.
  • Institute of Radiooncology-OncoRay, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany.
  • Department of Radiotherapy and Radiooncology, University Medical Center Giessen-Marburg, Marburg, Germany.
  • University of Applied Sciences, Institute of Medical Physics and Radiation Protection, Giessen, Germany.
  • Ion-Beam Therapy Center, Marburg, Germany.
  • Department of Radiation Oncology, Medical University of Vienna, Wien, Austria.

Förderung

  • European Metrology Programme for Innovation and Research

Zitieren

APA

Gebauer, B., Baumann, K. S., Fuchs, H., Georg, D., Oborn, B. M., Looe, H. K., & Lühr, A. (2024). Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane‐parallel ionization chamber. Medical Physics. https://doi.org/10.1002/mp.16797

Vancouver

Gebauer B, Baumann KS, Fuchs H, Georg D, Oborn BM, Looe HK, et al. Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane‐parallel ionization chamber. Medical Physics. 2024. doi:10.1002/mp.16797

BibTeX
@article{Gebauer2024,
  author = {Gebauer, Benjamin and Baumann, Kilian-Simon and Fuchs, Hermann and Georg, Dietmar and Oborn, Brad M and Looe, Hui-Khee and Lühr, Armin},
  title = {Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane‐parallel ionization chamber},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16797}
}
Kostenlos starten →