Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Lukas Birklein und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Computer Science, Johannes Gutenberg University, Mainz, Germany. Hauptthemen laut MEDLINE: Movement, Cone-Beam Computed Tomography. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Patient motions are a repeatedly reported phenomenon in oral and maxillofacial cone beam CT scans, leading to reconstructions of limited usability. In certain cases, independent movements of the mandible induce unpredictable motion patterns. Previous motion correction methods are not able to handle such complex cases of patient movements.

Purpose

Our goal was to design a combined motion estimation and motion correction approach for separate cranial and mandibular motions, solely based on the 2D projection images from a single scan.

Methods

Our iterative three‐step motion correction algorithm models the two articulated motions as independent rigid motions. First of all, we segment cranium and mandible in the projection images using a deep neural network. Next, we compute a 3D reconstruction with the poses of the object's trajectories fixed. Third, we improve all poses by minimizing the projection error while keeping the reconstruction fixed. Step two and three are repeated alternately.

Results

We find that our marker‐free approach delivers reconstructions of up to 85% higher quality, with respect to the projection error, and can improve on already existing techniques, which model only a single rigid motion. We show results of both synthetic and real data created in different scenarios. The reconstruction of motion parameters in a real environment was evaluated on acquisitions of a skull mounted on a hexapod, creating a realistic, easily reproducible motion profile.

Conclusions

The proposed algorithm consistently enhances the visual quality of motion impaired cone beam computed tomography scans, thus eliminating the need for a re‐scan in certain cases, considerably lowering radiation dosage for the patient. It can flexibly be used with differently sized regions of interest and is even applicable to local tomography.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MovementCone-Beam Computed TomographyHumansMotionSkullMandibleAlgorithmsImage Processing, Computer-AssistedPhantoms, ImagingArtifacts

Stichwörter der Autor:innen

cone beam computed tomographymotiontomography

Autor:innen und Institutionen

  • Institute of Computer Science, Johannes Gutenberg University, Mainz, Germany.
  • Computer Vision & Mixed Reality Group, RheinMain University of Applied Sciences, Wiesbaden, Germany.
  • Devision of Oral Diagnostic Sciences, Department of Oral Surgery and Stomatology, University of Bern, Bern, Switzerland.

Förderung

  • Deutsche Forschungsgemeinschaft

Zitieren

APA

Birklein, L., Niebler, S., Schömer, E., Brylka, R., Schwanecke, U., & Schulze, R. (2023). Motion correction for separate mandibular and cranial movements in cone beam CT reconstructions. Medical Physics. https://doi.org/10.1002/mp.16347

Vancouver

Birklein L, Niebler S, Schömer E, Brylka R, Schwanecke U, Schulze R. Motion correction for separate mandibular and cranial movements in cone beam CT reconstructions. Medical Physics. 2023. doi:10.1002/mp.16347

BibTeX
@article{Birklein2023,
  author = {Birklein, Lukas and Niebler, Stefan and Schömer, Elmar and Brylka, Robert and Schwanecke, Ulrich and Schulze, Ralf},
  title = {Motion correction for separate mandibular and cranial movements in cone beam CT reconstructions},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16347}
}
Kostenlos starten →