Technical note: Flat panel proton radiography with a patient specific imaging field for accurate WEPL assessment

Carmen Seller Oria, Jeffrey Free, Gabriel Guterres Marmitt, Barbara Knäusl, Sytze Brandenburg, Antje C Knopf, Arturs Meijers, Johannes A Langendijk, Stefan Both

Medical Physics · 2023

DOI: 10.1002/mp.16208 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Carmen Seller Oria und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. Hauptthemen laut MEDLINE: Protons, Proton Therapy. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: KWF Kankerbestrijding (Dutch Cancer Society): 11518.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Proton radiography (PR) uses highly energetic proton beams to create images where energy loss is the main contrast mechanism. Water‐equivalent path length (WEPL) measurements using flat panel PR (FP‐PR) have potential for in vivo range verification. However, an accurate WEPL measurement via FP‐PR requires irradiation with multiple energy layers, imposing high imaging doses.

Purpose

A FP‐PR method is proposed for accurate WEPL determination based on a patient‐specific imaging field with a reduced number of energies (n) to minimize imaging dose.

Methods

Patient‐specific FP‐PRs were simulated and measured for a head and neck (HN) phantom. An energy selection algorithm estimated spot‐wise the lowest energy required to cross the anatomy (Emin) using a water‐equivalent thickness map. Starting from Emin, n was restricted to certain values (n = 26, 24, 22, …, 2 for simulations, n = 10 for measurements), resulting in patient‐specific FP‐PRs. A reference FP‐PR with a complete set of energies was compared against patient‐specific FP‐PRs covering the whole anatomy via mean absolute WEPL differences (MAD), to evaluate the impact of the developed algorithm. WEPL accuracy of patient‐specific FP‐PRs was assessed using mean relative WEPL errors (MRE) with respect to measured multi‐layer ionization chamber PRs (MLIC‐PR) in the base of skull, brain, and neck regions.

Results

MADs ranged from 2.1 mm (n = 26) to 21.0 mm (n = 2) for simulated FP‐PRs, and 7.2 mm for measured FP‐PRs (n = 10). WEPL differences below 1 mm were observed across the whole anatomy, except at the phantom surfaces. Measured patient‐specific FP‐PRs showed good agreement against MLIC‐PRs, with MREs of 1.3 ± 2.0%, −0.1 ± 1.0%, and −0.1 ± 0.4% in the three regions of the phantom.

Conclusion

A method to obtain accurate WEPL measurements using FP‐PR with a reduced number of energies selected for the individual patient anatomy was established in silico and validated experimentally. Patient‐specific FP‐PRs could provide means of in vivo range verification.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ProtonsProton TherapyHumansWaterRadiographyPhantoms, ImagingHead

Stichwörter der Autor:innen

adaptive proton therapyflat panel detectorproton radiographyrange verificationwater equivalent path length

Substanzen

ProtonsWater

Autor:innen und Institutionen

  • Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Department of Radiation Oncology, Medical University of Vienna, Vienna, Austria.
  • Department of Internal Medicine, Center for Integrated Oncology Cologne, University Hospital of Cologne, Cologne, Germany.
  • Center for Proton Therapy, Paul Scherrer Institute, Villigen, Switzerland.

Förderung

  • KWF Kankerbestrijding (Dutch Cancer Society): 11518

Zitieren

APA

Seller Oria, C., Free, J., Marmitt, G. G., Knäusl, B., Brandenburg, S., Knopf, A. C., Meijers, A., Langendijk, J. A., & Both, S. (2023). Technical note: Flat panel proton radiography with a patient specific imaging field for accurate WEPL assessment. Medical Physics. https://doi.org/10.1002/mp.16208

Vancouver

Seller Oria C, Free J, Marmitt GG, Knäusl B, Brandenburg S, Knopf AC, et al. Technical note: Flat panel proton radiography with a patient specific imaging field for accurate WEPL assessment. Medical Physics. 2023. doi:10.1002/mp.16208

BibTeX
@article{SellerOria2023,
  author = {Seller Oria, Carmen and Free, Jeffrey and Marmitt, Gabriel Guterres and Knäusl, Barbara and Brandenburg, Sytze and Knopf, Antje C and Meijers, Arturs and Langendijk, Johannes A and Both, Stefan},
  title = {Technical note: Flat panel proton radiography with a patient specific imaging field for accurate WEPL assessment},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16208}
}
Kostenlos starten →