Camera‐based radiotherapy dosimetry using dual‐material 3D printed scintillator arrays

Nicholas Lynch, James L Robar, Thalat Monajemi

Medical Physics · 2023

DOI: 10.1002/mp.16167 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Nicholas Lynch und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Atmospheric Science, Dalhousie University, Halifax, NS, Canada. Hauptthemen laut MEDLINE: Radiometry, Software. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 33 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council of Canada, Beatrice Hunter Cancer Research Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose and objective To describe a methodology for the dual‐material fused deposition modeling (FDM) 3D printing of plastic scintillator arrays, to characterize their light output under irradiation using an sCMOS camera, and to establish a methodology for the dosimetric calibration of planar array geometries. Materials and methods We have published an investigation into the fabrication and characterization of single element FDM printed scintillators intending to produce customizable dosimeters for radiation therapy applications. 1 This work builds on previous investigations by extending the concept to the production of a high‐resolution (scintillating element size 3 × 3 × 3 mm 3 ) planar scintillator array. The array was fabricated using a BCN3D Epsilon W27 3D printer and composed of polylactic acid (PLA) filament and BCF‐10 plastic scintillator. The array's response was initially characterized using a 20 × 20 cm 2 6 MV photon field with a source‐to‐surface (SSD) distance of 100 cm and the beam incident on the top of the array. The light signals emitted under irradiation were imaged using 200 ms exposures from a sCMOS camera positioned at the foot of the treatment couch (210 cm from the array). The collected images were then processed using a purpose‐built software to correct known optical artefacts and determine the light output for each scintillating element. The light output was then corrected for element sensitivity and calibrated to dose using Monte Carlo simulations of the array and irradiation geometry based on the array's digital 3D print model. To assess the accuracy of the array calibration both a 3D beam and a clinical VMAT plan were delivered. Dose measurements using the calibrated array were then compared to EBT3 GAFChromic film and OSLD measurements, as well as Monte Carlo simulations and TPS calculations. Results Our results establish the feasibility of dual‐material 3D printing for the fabrication of custom plastic scintillator arrays. Assessment of the 3D printed scintillators response across each row of the array demonstrated a nonuniform response with an average percentage deviation from the mean of 2.1% ± 2.8%. This remains consistent with our previous work on individual 3D printed scintillators which showed an average difference of 2.3% and a maximum of 4.0% between identically printed scintillators. 1 Array dose measurements performed following calibration indicate difficulty in differentiating the scintillator response from ambient background light contamination at low doses (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiometrySoftwareRadiotherapy DosagePlasticsPrinting, Three-DimensionalMonte Carlo Method

Stichwörter der Autor:innen

3D printingdosimetryplastic scintillators

Substanzen

Plastics

Autor:innen und Institutionen

  • Department of Physics and Atmospheric Science, Dalhousie University, Halifax, NS, Canada.
  • Department of Medical Physics, Nova Scotia Health, Halifax, NS, Canada.
  • Department of Radiation Oncology, Dalhousie University, Halifax, NS, Canada.

Förderung

  • Natural Sciences and Engineering Research Council of Canada
  • Beatrice Hunter Cancer Research Institute

Zitieren

APA

Lynch, N., Robar, J. L., & Monajemi, T. (2023). Camera‐based radiotherapy dosimetry using dual‐material 3D printed scintillator arrays. Medical Physics. https://doi.org/10.1002/mp.16167

Vancouver

Lynch N, Robar JL, Monajemi T. Camera‐based radiotherapy dosimetry using dual‐material 3D printed scintillator arrays. Medical Physics. 2023. doi:10.1002/mp.16167

BibTeX
@article{Lynch2023,
  author = {Lynch, Nicholas and Robar, James L and Monajemi, Thalat},
  title = {Camera‐based radiotherapy dosimetry using dual‐material 3D printed scintillator arrays},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16167}
}
Kostenlos starten →