DIEP flap perfusion assessment using microdialysis versus Doppler ultrasonography: A comparative study
Microsurgery · 2024
DOI: 10.1002/micr.31097 ↗Kurz gefasst
Fachartikel in der Zeitschrift Microsurgery, 2024. Verfasst von Michail Sorotos und 5 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine and Psychology, Sapienza University of Rome - Sant'Andrea Hospital, Rome, Italy. Hauptthemen laut MEDLINE: Free Tissue Flaps, Mammaplasty. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
The increasing number of buried free‐tissue transfer procedures and the need for an objective method to evaluate vascular complications of free flaps has led to the development of new technologies. Microdialysis has been used to monitor free flaps using interstitial biological markers. Previous uses mainly focused on muscular flaps. Our aim is to compare external Doppler ultrasonography (EDU) evaluation versus microdialysis in the early follow‐up of adipocutaneous flaps, and propose an efficient postoperative monitoring protocol.
Methods
We retrospectively assessed 68 consecutive DIEP flaps (50 patients) performed between January 2019 and March 2021. All flaps received standardized post‐operative monitoring using clinical signs, EDU and microdialysis. Glucose and lactate concentrations were assessed using glucose 6 mmol/L as ischemic trend thresholds. We calculated Glucose/Lactate ratio as a new parameter for the assessment of flap viability.
Results
Among all the 68 flaps, two flaps returned to the operative theater when a combination of unsatisfactory microdialysis values and clinical/EDU signs identified vascular impairment; only one developed total flap necrosis. Reoperation rate was 2.94% with an overall flap success rate of 98.53%. External Doppler ultrasonography had 100% sensitivity and 82% specificity, while microdialysis had 100% sensitivity and 100% specificity.
Conclusions
Microdialysis values proved flap viability sooner than external Doppler ultrasonography, making it an excellent tool for post‐operative monitoring. With the appropriate thresholds for glucose and lactate concentrations, and glucose/lactate ratio used as a new parameter, it can help potentially avoiding unnecessary re‐explorations, and reducing flap ischemia times.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Faculty of Medicine and Psychology, Sapienza University of Rome - Sant'Andrea Hospital, Rome, Italy.
- Department of Human Pathology, Division and Residency Program in Plastic, Reconstructive and Aesthetic Surgery, University of Messina, Messina, Italy.
- Chair of Plastic Surgery, Faculty of Medicine and Psychology, Sapienza University of Rome - Sant'Andrea Hospital, Rome, Italy.
ORCID-Profile: Guido Firmani, Domenico Marrella, Fabio Santanelli di Pompeo
Zitieren
APA
Sorotos, M., Firmani, G., Tornambene, R., Marrella, D., Paolini, G., & Santanelli di Pompeo, F. (2024). DIEP flap perfusion assessment using microdialysis versus Doppler ultrasonography: A comparative study. Microsurgery. https://doi.org/10.1002/micr.31097
Vancouver
Sorotos M, Firmani G, Tornambene R, Marrella D, Paolini G, Santanelli di Pompeo F. DIEP flap perfusion assessment using microdialysis versus Doppler ultrasonography: A comparative study. Microsurgery. 2024. doi:10.1002/micr.31097
BibTeX
@article{Sorotos2024,
author = {Sorotos, Michail and Firmani, Guido and Tornambene, Roberta and Marrella, Domenico and Paolini, Guido and Santanelli di Pompeo, Fabio},
title = {DIEP flap perfusion assessment using microdialysis versus Doppler ultrasonography: A comparative study},
journal = {Microsurgery},
year = {2024},
doi = {10.1002/micr.31097}
}
EUVIMED