No Additional Benefit of 185 Hz Versus 130 Hz at Equivalent Energy in Deep Brain Stimulation for Tremor: A Randomized Prospective Clinical Trial
Movement Disorders Clinical Practice · 02.11.2026
DOI: 10.1002/mdc3.70886 ↗Kurz gefasst
Fachartikel in der Zeitschrift Movement Disorders Clinical Practice, online erschienen am 02.11.2026. Verfasst von Christina van der Linden und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurology, Faculty of Medicine and University Hospital Cologne, University of Cologne, Cologne, Germany. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Background
Deep brain stimulation (DBS) of the ventral intermediate nucleus and posterior subthalamic area (VIM/PSA) in Essential Tremor (ET) and the subthalamic nucleus (STN) in Parkinson's disease (PD) are established treatments for tremor. To achieve maximum tremor control, increasing stimulation frequency beyond 130 Hz is part of clinical practice, but lacks scientific evidence.
Objective
To compare tremor suppression under total electrical energy delivered (TEED)‐equivalent stimulation at 130 versus 185 Hz in VIM/PSA‐DBS for ET and STN‐DBS for PD.
Methods
In this prospective, double‐blind study, acute DBS effects were assessed in 18 people with ET ( n = 29 hemispheres), and 25 people with PD ( n = 30 hemispheres). Tremor‐suppressive effects, evaluated by accelerometry, were compared with TEED‐equivalent stimulation at 130 and 185 Hz using linear mixed‐effects models, explorative pairwise comparisons, and equivalency testing.
Results
Linear mixed‐effects models revealed no significant effect of stimulation frequency on tremor improvement in both cohorts. Pairwise comparisons showed no consistent differences in total tremor improvement with TEED‐equivalent 130 Hz vs 185 Hz DBS. Post‐hoc equivalence testing confirmed equivalence of stimulation frequencies under TEED‐equivalent conditions within a ±20% margin of relative tremor improvement.
Conclusion
This study provides Level II evidence that a higher stimulation frequency of 185 Hz does not offer additional benefit in deep brain stimulation for tremor in VIM/PSA for Essential Tremor or in the STN for Parkinson's Disease in the acute setting and supports 130 Hz as the standard stimulation frequency for tremor suppression.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurology, Faculty of Medicine and University Hospital Cologne, University of Cologne, Cologne, Germany.
- Department of Neurology with Experimental Neurology, Charité-Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt Universität zu Berlin, Berlin, Germany.
- Cognitive Neuroscience, Institute of Neuroscience and Medicine (INM3)-Research Center Jülich, Jülich, Germany.
- Department of Stereotactic and Functional Neurosurgery, Faculty of Medicine and University Hospital Cologne, University of Cologne, Cologne, Germany.
ORCID-Profile: Christina van der Linden, Peer Trapp, Till A Dembek, Charlotte Schedlich-Teufer, Gregor A Brandt, Gereon R Fink, Veerle Visser-Vandewalle, Michael T Barbe, Jan Niklas Petry-Schmelzer
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
van der Linden, C., Trapp, P., Dembek, T. A., Schedlich-Teufer, C., Brandt, G. A., Jergas, H., Fink, G. R., Visser-Vandewalle, V., Barbe, M. T., & Petry-Schmelzer, J. N. (2026). No Additional Benefit of 185 Hz Versus 130 Hz at Equivalent Energy in Deep Brain Stimulation for Tremor: A Randomized Prospective Clinical Trial. Movement Disorders Clinical Practice. https://doi.org/10.1002/mdc3.70886
Vancouver
van der Linden C, Trapp P, Dembek TA, Schedlich-Teufer C, Brandt GA, Jergas H, et al. No Additional Benefit of 185 Hz Versus 130 Hz at Equivalent Energy in Deep Brain Stimulation for Tremor: A Randomized Prospective Clinical Trial. Movement Disorders Clinical Practice. 2026. doi:10.1002/mdc3.70886
BibTeX
@article{vanderLinden2026,
author = {van der Linden, Christina and Trapp, Peer and Dembek, Till A and Schedlich-Teufer, Charlotte and Brandt, Gregor A and Jergas, Hannah and Fink, Gereon R and Visser-Vandewalle, Veerle and Barbe, Michael T and Petry-Schmelzer, Jan Niklas},
title = {No Additional Benefit of 185 Hz Versus 130 Hz at Equivalent Energy in Deep Brain Stimulation for Tremor: A Randomized Prospective Clinical Trial},
journal = {Movement Disorders Clinical Practice},
year = {2026},
doi = {10.1002/mdc3.70886}
}