Fachartikel in der Zeitschrift Lasers in Surgery and Medicine, 2025. Verfasst von Aditi Maheshwari und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dermatology, Harvard Medical School, Massachusetts General Hospital, Cutaneous Biology Research Center, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Laser Therapy. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
Fractional Photothermolysis (FP) relies on precise laser focus to create microscopic treatment zones (MTZs). Focal plane deviations, caused by handpiece pressure and movement, shift the treated tissue out of focus, increasing spot size and reducing lesion depth, compromising treatment efficacy. This study investigates an optimized spacer to mitigate load‐induced deviations and improve clinical outcomes.
Methods
A difference frequency generation (DFG) laser system (IPG Medical) with 3050/3200 nm wavelength treated ex vivo human abdominoplasty skin, mounted on a load‐adjustable platform (0–1000 g, ±0.1 g). Tissue was treated at 5 mJ/pulse with a nominal spot size of 40 µm using spacers ranging from 2 × 2 mm 2 to 8 × 8 mm 2 and an optimized 2 × 2 mm 2 grid spacer. Optical coherence tomography measured skin bulging and focal plane deviations, while histological analysis validated lesion depth and diameter. Knife‐edge technique projections were used to assess theoretical beam behavior.
Results
Applied load caused skin bulging, resulting in a linear increase in lesion diameter and a linear decrease in depth. Smaller spacers (2 × 2 mm 2 ) reduced focal deviations and maintained MTZ precision within the Rayleigh range. 2 × 2 mm 2 grid spacers further reduced bulging by redistributing the load and ensuring more uniform MTZ formation.
Conclusion
Load‐induced focal shifts significantly impact FP efficiency, particularly on nonuniform skin surfaces. An optimized spacer enhanced precision by minimizing bulging, allowing treatments to remain within one Rayleigh length. These findings highlight the clinical importance of spacer selection in achieving consistent FP outcomes, particularly for small‐spot laser applications. Clinicians should integrate smaller or grid‐patterned spacers to enhance procedural accuracy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Dermatology, Harvard Medical School, Massachusetts General Hospital, Cutaneous Biology Research Center, Boston, Massachusetts, USA.
ORCID-Profile: Michael Wang-Evers
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Maheshwari, A., Manstein, N., Warner-Levy, J., Warner-Levy, A., Alfonso, C. M., Doyle, A. E., & Wang-Evers, M. (2025). Optimized Distance Holders Improve Precision in Fractional Laser Treatment. Lasers in Surgery and Medicine. https://doi.org/10.1002/lsm.70050
Vancouver
Maheshwari A, Manstein N, Warner-Levy J, Warner-Levy A, Alfonso CM, Doyle AE, et al. Optimized Distance Holders Improve Precision in Fractional Laser Treatment. Lasers in Surgery and Medicine. 2025. doi:10.1002/lsm.70050
BibTeX
@article{Maheshwari2025,
author = {Maheshwari, Aditi and Manstein, Nelson and Warner-Levy, John and Warner-Levy, Alexander and Alfonso, Connor M and Doyle, Abigail E and Wang-Evers, Michael},
title = {Optimized Distance Holders Improve Precision in Fractional Laser Treatment},
journal = {Lasers in Surgery and Medicine},
year = {2025},
doi = {10.1002/lsm.70050}
}
EUVIMED