Development of a Novel Endotracheal Tube to Measure Posterior Laryngeal Pressure

Denis Lafreniere iD, Amelia Geist

The Laryngoscope · 2026

DOI: 10.1002/lary.70113 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Laryngoscope, 2026. Verfasst von Denis Lafreniere und 1 weiteren Autor:innen. Beteiligte Einrichtung: Division of Otolaryngology, UConn Health, Farmington, Connecticut, USA. Hauptthemen laut MEDLINE: Intubation, Intratracheal, Larynx, Glottis. Die Arbeit nennt 22 Literaturangaben. Gefördert von: University of Connecticut.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The development of an endotracheal tube (ETT) prototype to measure the pressure on the posterior glottis during intubation.

Methods

Three ETTs , size 6, 7, and 8 mm, were modified with a piezoelectric pressure sensor. These ETTs were then used to intubate a laryngeal model. Pressures were measured at a hypopharyngeal laryngeal angle ( HLA ) of 90°, and the laryngeal platform was raised to create a more acute HLA ranging from 90° to 45°. These same ETTs were also used to intubate the larynx of a fresh frozen cadaver, with pressure measured with the head in neutral and flexed positions.

Results

The pressures exerted by the 6 mm ETT ranged from 7.33 cm H 2 O at 90 degrees HLA to 62.57 cm H 2 O at 45 degrees HLA . The 7 mm ETT ranged from 5.42 to 451.19 cm H 2 O and the 8 mm ETT pressure ranged from 136.98 to 534.38 cm H 2 O , respectively. The differences were statistically significant. The cadaveric study revealed the average pressure for the 6‐, 7‐, and 8‐mm ETT to be 2.03, 166.78, and 228.49 cm H 2 O , respectively, in the neutral position and 0.85, 30.37, and 158.51 cm H 2 O , respectively, in the flexed position. These differences were statistically significant.

Conclusion

Posterior laryngeal pressure monitoring can be conducted utilizing ETT‐based pressure sensors. Progression of this technology may help minimize laryngeal injury caused by intubation by determining which patients may be optimal candidates for early tracheostomy. Level of Evidence 5

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intubation, IntratrachealLarynxGlottisHumansPressureCadaverEquipment Design

Stichwörter der Autor:innen

endotracheal tubeposterior glottic stenosispressure sensor

Autor:innen und Institutionen

  • Division of Otolaryngology, UConn Health, Farmington, Connecticut, USA.

ORCID-Profile: Denis Lafreniere

Förderung

  • University of Connecticut

Zitieren

APA

Lafreniere, D., & Geist, A. (2026). Development of a Novel Endotracheal Tube to Measure Posterior Laryngeal Pressure. The Laryngoscope. https://doi.org/10.1002/lary.70113

Vancouver

Lafreniere D, Geist A. Development of a Novel Endotracheal Tube to Measure Posterior Laryngeal Pressure. The Laryngoscope. 2026. doi:10.1002/lary.70113

BibTeX
@article{Lafreniere2026,
  author = {Lafreniere, Denis and Geist, Amelia},
  title = {Development of a Novel Endotracheal Tube to Measure Posterior Laryngeal Pressure},
  journal = {The Laryngoscope},
  year = {2026},
  doi = {10.1002/lary.70113}
}
Kostenlos starten →