Fachartikel in der Zeitschrift The Laryngoscope, 2025. Verfasst von Joseph M Chen und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada. Hauptthemen laut MEDLINE: Cochlear Implantation, Synchrotrons, Temporal Bone, Robotic Surgical Procedures. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Canada Foundation for Innovation (CFI), Natural Sciences and Engineering Research Council (NSERC), National Research Council (NRC).
Abstract (Originalsprache)
Objectives
Minimizing intraoperative trauma during cochlear implant electrode insertion is crucial for preserving residual hearing. Manual insertions are subject to variability due to surgeon experience and anatomical differences, often leading to inconsistent outcomes. Robotic‐assisted systems have been developed to improve insertion precision and reduce trauma. This study evaluates the safety and efficacy of the OTODRIVE system for cochlear implant electrode insertion.
Methods
Fifteen cadaveric human temporal bones were implanted with a variety of electrodes (FLEX26, FLEX28, FLEXSOFT, FLEX34) and an inner ear catheter (INCAT). The OTODRIVE system was used to maintain a constant insertion speed of 0.1 mm/s. The samples were subsequently imaged using synchrotron radiation phase‐contrast imaging (SR‐PCI) to assess intracochlear trauma. Parameters such as linear and angular insertion depths and cochlear coverage were measured. The presence of trauma was analyzed using a trauma evaluation plot.
Results
The average linear insertion depth was 24.8 ± 3.5 mm, and the average angular insertion depth was 527° ± 81°. Cochlear coverage was 74% ± 7%. Only one case of partial scala media translocation and two cases of basilar membrane elevation were observed, with no instances of severe trauma such as tip fold‐over, scala vestibuli translocation, or osseous spiral lamina fracture.
Conclusion
The findings indicate that the OTODRIVE system enables consistent and controlled electrode placement while minimizing trauma to cochlear structures, suggesting its potential to enhance hearing and structural preservation outcomes. Further clinical studies are required to establish the long‐term benefits of robotic‐assisted techniques over manual insertions in routine cochlear implantation. Level of Evidence N/A.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.
- Sunnybrook Research Institute, University of Toronto, Toronto, Canada.
- Department of Otorhinolaryngology, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
- Department of Otolaryngology-Head and Neck Surgery, Western University London, London, Ontario, Canada.
- Department of Medical Biophysics and Department of Electrical and Computer Engineering, Western University London, London, Ontario, Canada.
ORCID-Profile: Vincent Y W Lin, Jennifer L Spiegel, Omer J Ungar
Förderung
- Canada Foundation for Innovation (CFI)
- Natural Sciences and Engineering Research Council (NSERC)
- National Research Council (NRC)
- Canadian Institutes of Health Research (CIHR)
- Government of Saskatchewan, and the University of Saskatchewan
- Natural Sciences and Engineering Research Council of Canada
- National Research Council Canada
- Government of Saskatchewan
Zitieren
APA
Chen, J. M., Lin, V. Y. W., Le, T. N., Spiegel, J. L., Ungar, O. J., Bajin, M. D., Ladak, H. M., & Agrawal, S. (2025). Synchrotron‐Based Trauma Assessment of Robotic Electrode Insertions in Cochlear Implantation. The Laryngoscope. https://doi.org/10.1002/lary.32254
Vancouver
Chen JM, Lin VYW, Le TN, Spiegel JL, Ungar OJ, Bajin MD, et al. Synchrotron‐Based Trauma Assessment of Robotic Electrode Insertions in Cochlear Implantation. The Laryngoscope. 2025. doi:10.1002/lary.32254
BibTeX
@article{Chen2025,
author = {Chen, Joseph M and Lin, Vincent Y W and Le, Trung N and Spiegel, Jennifer L and Ungar, Omer J and Bajin, Munir D and Ladak, Hanif M and Agrawal, Sumit},
title = {Synchrotron‐Based Trauma Assessment of Robotic Electrode Insertions in Cochlear Implantation},
journal = {The Laryngoscope},
year = {2025},
doi = {10.1002/lary.32254}
}
EUVIMED