EUVIMEDEUVIMED entdecken
Crossref

Feasibility of Obtaining Intraoperative Electrically Evoked Stapedial Reflex Thresholds ( eSRT )

Isabelle J Chau iD, Eric Cunningham, Rivka Bornstein, Regina Albinus, Katrina R Stidham

The Laryngoscope · 2025

DOI: 10.1002/lary.32226 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Laryngoscope, 2025. Verfasst von Isabelle J Chau und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Medicine, New York Medical College, Valhalla, New York, USA. Hauptthemen laut MEDLINE: Cochlear Implantation, Cochlear Implants, Monitoring, Intraoperative, Reflex, Acoustic. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 48 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To assess the rate of obtaining intraoperative eSRT measurements in primary cochlear implant (CI) recipients while limiting the use of volatile anesthetics, such as sevoflurane.

Methods

In this prospective cohort study, thirty‐one patients who underwent CI surgery with a Cochlear Americas device from January 2023 through March 2024 were included. Intraoperative eSRT testing at electrodes E1, E6, E12, E17, and E22 was performed on all patients. Sevoflurane was discontinued and confirmed undetectable at least 10 min prior to eSRT testing to avoid diminishing the stapedial reflex. The primary outcome measure was eSRT response rate at any electrode or pulse width.

Results

Thirty‐four CIs were placed in 31 patients, including three bilateral recipients. Eighty‐five percent (29/34) of implants had at least one response at any electrode or pulse width; 71% (24/34) at pulse width 25 μs; and 82% (28/34) at 25 or 37 μs. Response rates were significantly higher at apical (E17, E22) and middle (E12) electrodes compared to basal electrodes (E1, E6) ( p = 0.004). Comparisons also revealed significantly different eSRT values between apical, middle, and basal electrodes ( p < 0.0001). Specifically, the basal electrode region had a greater median threshold (30.5 nC/phase) than apical (18.3 nC/phase; p < 0.0001) and middle (17.7 nC/phase; p < 0.001) regions.

Conclusion

Intraoperative eSRT is a feasible and effective way of obtaining an objective measurement that can potentially guide CI programming. Subsequent studies will investigate the relationship between intraoperative and postoperative eSRT in this patient cohort and determine how intraoperative eSRT may facilitate CI programming. Level of Evidence 2.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cochlear ImplantationCochlear ImplantsMonitoring, IntraoperativeReflex, AcousticStapediusHumansFemaleProspective StudiesMaleMiddle AgedFeasibility StudiesAgedAdultSevofluraneAuditory ThresholdElectric Stimulation

Stichwörter der Autor:innen

cochlear implanteSRTelectrically evoked stapedial reflex thresholdelectrode stimulationfittingstapedial reflex

Substanzen

Sevoflurane

Autor:innen und Institutionen

  • School of Medicine, New York Medical College, Valhalla, New York, USA.
  • Department of Otolaryngology-Head & Neck Surgery, Westchester Medical Center, Valhalla, New York, USA.

ORCID-Profile: Isabelle J Chau

Zitieren

APA

Chau, I. J., Cunningham, E., Bornstein, R., Albinus, R., & Stidham, K. R. (2025). Feasibility of Obtaining Intraoperative Electrically Evoked Stapedial Reflex Thresholds ( eSRT ). The Laryngoscope. https://doi.org/10.1002/lary.32226

Vancouver

Chau IJ, Cunningham E, Bornstein R, Albinus R, Stidham KR. Feasibility of Obtaining Intraoperative Electrically Evoked Stapedial Reflex Thresholds ( eSRT ). The Laryngoscope. 2025. doi:10.1002/lary.32226

BibTeX
@article{Chau2025,
  author = {Chau, Isabelle J and Cunningham, Eric and Bornstein, Rivka and Albinus, Regina and Stidham, Katrina R},
  title = {Feasibility of Obtaining Intraoperative Electrically Evoked Stapedial Reflex Thresholds ( eSRT )},
  journal = {The Laryngoscope},
  year = {2025},
  doi = {10.1002/lary.32226}
}
Kostenlos starten →