EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Does the TruBlue Laser Set Microlaryngoscopy Equipment on Fire? A Systematic Evaluation

Ariel Roitman iD, Tadeas Lunga iD, Anumitha Venkatraman iD, Kristopher M Schroeder, Susan L Thibeault, Seth H Dailey iD

The Laryngoscope · 2025

DOI: 10.1002/lary.31943 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Laryngoscope, 2025. Verfasst von Ariel Roitman und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, Division of Otolaryngology-Head and Neck Surgery, School of Medicine and Public Health, University of Wisconsin-Madison, Madis…. Hauptthemen laut MEDLINE: Laryngoscopy, Fires, Laser Therapy, Laryngoscopes. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The risk of fire during laser microlaryngoscopy is well known. However, limited information is available about fire risk with the novel TruBlue laser. This study systematically evaluates its interactions with common surgical supplies, offering valuable insights into safety considerations for surgeons. Material and Methods We used experimental conditions to test the extent to which TruBlue laser energy produces smoke, perforation, or fire in Rüsch®, Medtronic™ and microlaryngeal endotracheal tubes and in surgical pledgets.

Results

Only the Microlaryngeal Tube (MLT) caught fire. Notably, it happened only when the laser fiber shifted on the tube's surface. Smoke emerged solely from the laser fiber applied to the Medtronic™ shaft and only during continuous contact mode. Cuff perforation and smoke emanating from the shaft occurred in three‐quarters of the Rüsch® trials. The pledgets' radiopaque segment exhibited a greater combustibility than other segments ( p < 0.01). In many of the pledget trials, faster smoke emission occurred with shorter laser‐to‐target distances ( p < 0.05). Water‐soaked pledgets displayed a reduced rate of smoke production ( p < 0.01) and string division.

Conclusion

The Medtronic™ tube assures remarkable safety with a nonignitable shaft and low cuff ignition. The MLT poses the highest ignition risk. Cuff perforation risk is mitigated by maintaining a 0.3‐cm distance from the laser fiber tip. Pledget fire risk is mitigated by positioning the radiopaque part away from the laser beam and by soaking the pledget with water. Laser division of the pledgets' string was common. Level of Evidence NA Laryngoscope , 135:1752–1758, 2025

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaryngoscopyFiresLaser TherapyLaryngoscopesHumansEquipment DesignSmoke

Stichwörter der Autor:innen

TruBlue laserairway surgeryexperimental studyfire riskmicrolaryngoscopysurgical supplies

Substanzen

Smoke

Autor:innen und Institutionen

  • Department of Surgery, Division of Otolaryngology-Head and Neck Surgery, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, U.S.A.
  • Ruth and Bruce Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Haifa, Israel.
  • Department of Anesthesiology, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, U.S.A.

ORCID-Profile: Ariel Roitman, Tadeas Lunga, Anumitha Venkatraman, Seth H Dailey

Zitieren

APA

Roitman, A., Lunga, T., Venkatraman, A., Schroeder, K. M., Thibeault, S. L., & Dailey, S. H. (2025). Does the TruBlue Laser Set Microlaryngoscopy Equipment on Fire? A Systematic Evaluation. The Laryngoscope. https://doi.org/10.1002/lary.31943

Vancouver

Roitman A, Lunga T, Venkatraman A, Schroeder KM, Thibeault SL, Dailey SH. Does the TruBlue Laser Set Microlaryngoscopy Equipment on Fire? A Systematic Evaluation. The Laryngoscope. 2025. doi:10.1002/lary.31943

BibTeX
@article{Roitman2025,
  author = {Roitman, Ariel and Lunga, Tadeas and Venkatraman, Anumitha and Schroeder, Kristopher M and Thibeault, Susan L and Dailey, Seth H},
  title = {Does the TruBlue Laser Set Microlaryngoscopy Equipment on Fire? A Systematic Evaluation},
  journal = {The Laryngoscope},
  year = {2025},
  doi = {10.1002/lary.31943}
}
Kostenlos starten →