Validating an Evoked Potential Platform for Electrocochleography During Cochlear Implantation
The Laryngoscope · 2024
DOI: 10.1002/lary.31724 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift The Laryngoscope, 2024. Verfasst von Jordan J Varghese und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology - Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, U.S.A. Hauptthemen laut MEDLINE: Cochlear Implantation, Audiometry, Evoked Response, Evoked Potentials, Auditory. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDCD NIH HHS, NCATS NIH HHS, American Hearing Research Foundation.
Abstract (Originalsprache)
Objective
To validate electrocochleography (ECochG) between an auditory evoked potential (AEP) machine and an established cochlear implant (CI) manufacturer ECochG system.
Methods
Intraoperative validation study at a tertiary referral center. Patients included adults and children undergoing cochlear implantation. Intraoperative ECochG was measured with both the Intelligent Hearing Systems (IHS) Duet AEP machine and Cochlear Corporation (CC) ECochG platform. Recording electrodes captured extracochlear measurements through a standard facial recess. Tone‐bursts were presented from 250 Hz to 2 kHz (~110 dB SPL). A fast Fourier transform (FFT) of ECochG waveforms at key frequencies was summed into a total response (ECochG‐TR). Pearson's correlation was utilized to evaluate the relationship between IHS‐ECochG‐TR and CC‐ECochG‐TR after confirming normality.
Results
Thirty patients were enrolled with an average age of 67 years (SD 18.8). In the ear that was implanted, mean preoperative pure‐tone average (PTA; 0.5, 1, 2, and 4 kHz) was 87.4 dB HL (SD 19.3) and mean preoperative word‐recognition scores (WRS) was 17.0% correct (SD 19.1). There was strong correlation ( r = 0.905, 95% confidence interval: 0.809 to 0.954) between IHS‐ECochG‐TR (median 2.30 μV, range 0.1–148.26) and CC‐ECochG‐TR (median 3.00 μV, range 0.1–239.63). Four patients underwent transtympanic ECochG with the IHS system for feasibility evaluation and achieved similar responses.
Conclusion
Extracochlear ECochG has been predictive of CI speech perception performance. The IHS duet system is a valid measure of extracochlear ECochG for the CI population. Future work will utilize this system for measuring transtympanic ECochG to improve preoperative estimation of CI performance. Level of Evidence 3 Laryngoscope , 2024
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology - Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, U.S.A.
ORCID-Profile: Jordan J Varghese, Matthew A Shew, Amit Walia, Nedim Durakovic, Jacques A Herzog
Förderung
- NIDCD NIH HHS
- NCATS NIH HHS
- American Hearing Research Foundation
- National Institute on Deafness and Other Communication Disorders
- National Center for Advancing Translational Sciences
Zitieren
APA
Varghese, J. J., Shew, M. A., Walia, A., Lefler, S. M., Durakovic, N., Wick, C. C., Ortmann, A. J., Herzog, J. A., & Buchman, C. A. (2024). Validating an Evoked Potential Platform for Electrocochleography During Cochlear Implantation. The Laryngoscope. https://doi.org/10.1002/lary.31724
Vancouver
Varghese JJ, Shew MA, Walia A, Lefler SM, Durakovic N, Wick CC, et al. Validating an Evoked Potential Platform for Electrocochleography During Cochlear Implantation. The Laryngoscope. 2024. doi:10.1002/lary.31724
BibTeX
@article{Varghese2024,
author = {Varghese, Jordan J and Shew, Matthew A and Walia, Amit and Lefler, Shannon M and Durakovic, Nedim and Wick, Cameron C and Ortmann, Amanda J and Herzog, Jacques A and Buchman, Craig A},
title = {Validating an Evoked Potential Platform for Electrocochleography During Cochlear Implantation},
journal = {The Laryngoscope},
year = {2024},
doi = {10.1002/lary.31724}
}
EUVIMED