Long‐Term Plate Complications in Patient‐Specific Plates Utilizing Computer‐Aided Design
The Laryngoscope · 2024
DOI: 10.1002/lary.31640 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Laryngoscope, 2024. Verfasst von Sara Yang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, Oregon Health and Science University, Portland, Oregon, U.S.A. Hauptthemen laut MEDLINE: Bone Plates, Computer-Aided Design, Postoperative Complications, Free Tissue Flaps. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben.
Abstract (Originalsprache)
Objective
Assess the long‐term plate complications with patient‐specific plates (PSPs) created with computer‐aided design (CAD) and computer‐aided manufacturing (CAM) for fibula free flap reconstructions for mandibular defects.
Methods
Retrospective chart review from January 2010 to July 2022 of patients who underwent mandibular reconstruction with a fibula free flap and PSP. Primary outcome was plating‐related complications, defined as plate exposure, fracture, loose screws, and plate removal.
Results
A total of 221 patients underwent PSP fibula reconstruction. Average age was 59.8 + 14.3 years old with male to female ratio of 2:1. Squamous cell carcinoma of the mandible was the most common reason for resection, 47.5%, n = 105. Plate removal occurred in 11% of patients ( n = 25) about 17.4 months after the initial surgery. Plates were removed due to exposure (76%, n = 19) or screw loosening (24%, n = 6). Malignancy was associated with an increased risk of plate complications when compared to benign tumor (odds ratio [OR] 9.04, confidence interval [CI] 1.36–3.85), osteonecrosis (OR 1.38, CI 0.59–3.48), and trauma (OR 1.26, CI 0.23–12.8). Postoperative radiation therapy (OR 2.27, CI 1.07–4.82, p = 0.026) and surgical site infection (OR 9.22, CI 4.11–21.88, p = 0.001) were associated with more plate complications.
Conclusions
CAD creates PSPs that remain stable in the majority of patients over the long term. Plate removal is less compared to non‐PSP reconstruction. Consideration of the soft tissue envelope over the plate and management of perioperative infection at the time of surgery should be entertained. Level of Evidence 4 Laryngoscope , 134:4929–4934, 2024
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, Oregon Health and Science University, Portland, Oregon, U.S.A.
- Department of Otolaryngology-Head and Neck Surgery, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, U.S.A.
- Department of Oral Maxillofacial Surgery, Oregon Health Science University, Portland, Oregon, U.S.A.
- Oregon Health Science University School of Medicine, Portland, Oregon, U.S.A.
ORCID-Profile: Sara Yang, Gavin Young
Zitieren
APA
Yang, S., Morton, Z., Sautter, M., Young, G., Petrisor, D., Chandra, S. R., & Wax, M. K. (2024). Long‐Term Plate Complications in Patient‐Specific Plates Utilizing Computer‐Aided Design. The Laryngoscope. https://doi.org/10.1002/lary.31640
Vancouver
Yang S, Morton Z, Sautter M, Young G, Petrisor D, Chandra SR, et al. Long‐Term Plate Complications in Patient‐Specific Plates Utilizing Computer‐Aided Design. The Laryngoscope. 2024. doi:10.1002/lary.31640
BibTeX
@article{Yang2024,
author = {Yang, Sara and Morton, Zoey and Sautter, Mackenzie and Young, Gavin and Petrisor, Daniel and Chandra, Srinivasa R and Wax, Mark K},
title = {Long‐Term Plate Complications in Patient‐Specific Plates Utilizing Computer‐Aided Design},
journal = {The Laryngoscope},
year = {2024},
doi = {10.1002/lary.31640}
}
EUVIMED