Comparison of Angiolytic Effects Between the 445‐nm Blue Laser and 532‐nm Pulsed KTP Laser

Duy Duong Nguyen iD, Jing-Yin Pang, Daniel Novakovic iD

The Laryngoscope · 2024

DOI: 10.1002/lary.31364 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Laryngoscope, 2024. Verfasst von Duy Duong Nguyen und 2 weiteren Autor:innen. Beteiligte Einrichtung: Voice Research Laboratory, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Lasers, Solid-State, Chorioallantoic Membrane. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Dr Liang Voice Program, The University of Sydney.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to compare the selective absorption of the 445‐nm Blue laser (BL) and the 532‐nm pulsed potassium‐titanyl‐phosphate (KTP) laser by blood vessels.

Methods

Thirty‐six chicken eggs at day 14 of incubation were dissected to expose the chick chorioallantoic membrane (CAM). Third‐order vessels of the CAM were identified and irradiated using BL and KTP lasers using various settings at a laser‐to‐vessel distance of 3 mm using 0.4 mm fiber size. In total, 494 vessels segments were irradiated. Mean (standard deviation) number of irradiations for each setting was 26.0 (4.6), range from 15 to 39. Outcome measures included ablation rate (AR) and rupture rate (RR).

Results

The two lasers were compared for AR and RR at long and medium pulse width (PW) associated with different power levels. At long PW (above 100 ms), BL showed significantly higher AR than KTP at high energy (600 mJ/pulse) and low energy (400 mJ/pulse); they did not show different AR and RR at medium energy levels (500 mJ/pulse). Using medium PW settings plus high and medium energy levels, BL and KTP showed relatively high AR and did not significantly differ in performance. However, at medium PW plus low energy (400–450 mJ/pulse), KTP showed significantly higher AR compared to BL.

Conclusion

At long PW, BL appeared to show higher AR than KTP at high or low energy levels, but they showed equivalent performance at medium energy. At medium PW, both performed similarly from high to medium energy, but KTP appeared to perform better than BL at lower energy settings. Level of Evidence NA Laryngoscope , 134:3220–3225, 2024

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lasers, Solid-StateChorioallantoic MembraneAnimalsChick EmbryoBlood Vessels

Stichwörter der Autor:innen

445‐nm laser532‐nm laserKTP laserblue lasermicrolaryngeal surgeryphotoangiolysis

Autor:innen und Institutionen

  • Voice Research Laboratory, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
  • National Hospital of Otorhinolaryngology, Hanoi, Vietnam.
  • Department of Ear-Nose-Throat, Khoo Teck Puat Hospital, Yishun Central, Singapore.
  • The Canterbury Hospital, Sydney, New South Wales, Australia.

ORCID-Profile: Duy Duong Nguyen, Daniel Novakovic

Förderung

  • Dr Liang Voice Program, The University of Sydney

Zitieren

APA

Nguyen, D. D., Pang, J. Y., & Novakovic, D. (2024). Comparison of Angiolytic Effects Between the 445‐nm Blue Laser and 532‐nm Pulsed KTP Laser. The Laryngoscope. https://doi.org/10.1002/lary.31364

Vancouver

Nguyen DD, Pang JY, Novakovic D. Comparison of Angiolytic Effects Between the 445‐nm Blue Laser and 532‐nm Pulsed KTP Laser. The Laryngoscope. 2024. doi:10.1002/lary.31364

BibTeX
@article{Nguyen2024,
  author = {Nguyen, Duy Duong and Pang, Jing-Yin and Novakovic, Daniel},
  title = {Comparison of Angiolytic Effects Between the 445‐nm Blue Laser and 532‐nm Pulsed KTP Laser},
  journal = {The Laryngoscope},
  year = {2024},
  doi = {10.1002/lary.31364}
}
Kostenlos starten →