A novel knee implant for total knee arthroplasty meets expectations at 10 years. First long‐term follow‐up report of clinical outcomes and survivorship
Knee Surgery, Sports Traumatology, Arthroscopy · 2026-04 · 34 (4) · 1318-1326
DOI: 10.1002/ksa.70037 ↗Kurz gefasst
Fachartikel in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, online erschienen am 29.08.2025. Verfasst von Alice Montagna und 6 weiteren Autor:innen. Beteiligte Einrichtung: U.O.C Ortopedia e Traumatologia Fondazione Poliambulanza, Brescia, Italy. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Knee Prosthesis, Osteoarthritis, Knee. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Achieving a “forgotten knee” after total knee arthroplasty (TKA) remains a primary goal in modern knee replacement surgery. Anatomic implant designs aim to replicate native knee anatomy and kinematics, potentially improving patient satisfaction and functional outcomes. This study evaluates the long‐term clinical outcome and survivorship of the Persona Knee System at a minimum follow‐up of 10 years.
Methods
116 TKAs performed using the Persona Posterior Stabilised (PS) Knee System (Zimmer Biomet, Warsaw, Indiana, USA) between 2013 and 2014 at a high‐volume orthopaedic centre were prospectively followed and retrospectively analysed. Clinical outcomes were assessed using patient‐reported outcome measures (PROMs), including the Forgotten Joint Score‐12 (FJS‐12), Oxford Knee Score (OKS), and Western Ontario and McMaster Universities Arthritis Index (WOMAC). Implant survivorship was determined using Kaplan‐Meier analysis, and complication rates were recorded.
Results
At a mean follow‐up of 11.1 years, 116 knees were available for analysis. Patients were assessed clinically and radiographically at 1, 3 and 6 months postoperatively, and then annually, with a mean follow‐up of 11.1 years. The mean FJS‐12 was 69.52 (SD 12.21, range 15–88), indicating a high level of joint awareness reduction. The OKS and WOMAC scores significantly improved postoperatively, with mean final values of 38.63 (SD 7.99, range 7–48) and 25.29 (SD 16.97, range 4–91), respectively. Radiological analysis demonstrated accurate and stable implant positioning, with no progressive radiolucent lines in non‐revised cases. Kaplan–Meier survival analysis showed a 95.7% (SD 1.9%) implant survival rate. The revision rate was 4.3%, with aseptic loosening and persistent painful prosthesis as the primary causes.
Conclusion
The anatomic design of the Persona Knee System provides excellent long‐term clinical outcomes, high patient satisfaction, and sustained implant durability. Future research should further investigate patient‐specific factors and surgical refinements to optimise long‐term outcomes in TKA. Level of Evidence Level III.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- U.O.C Ortopedia e Traumatologia Fondazione Poliambulanza, Brescia, Italy.
- Clinica Ortopedica e Traumatologica Fondazione IRCCS Policlinico San Matteo, Pavia, Italy.
- Università Cattolica del Sacro Cuore, Rome, Italy.
- IUSS Istituto Universitario di Studi Superiori, Pavia, Italy.
- Unità di Chirurgia Robotica U.O.C Ortopedia e Traumatologia, Fondazione Poliambulanza, Brescia, Italy.
- Department of Life Science, Health, and Health Professions Università degli Studi "Link Campus University", Roma, Italy.
ORCID-Profile: Stefano Marco Paolo Rossi
Zitieren
APA
Montagna, A., Marescalchi, M., Cinelli, V., Sangaletti, R., Andriollo, L., Benazzo, F., & Rossi, S. M. P. (2026). A novel knee implant for total knee arthroplasty meets expectations at 10 years. First long‐term follow‐up report of clinical outcomes and survivorship. Knee Surgery, Sports Traumatology, Arthroscopy, 34(4), 1318-1326. https://doi.org/10.1002/ksa.70037
Vancouver
Montagna A, Marescalchi M, Cinelli V, Sangaletti R, Andriollo L, Benazzo F, et al. A novel knee implant for total knee arthroplasty meets expectations at 10 years. First long‐term follow‐up report of clinical outcomes and survivorship. Knee Surgery, Sports Traumatology, Arthroscopy. 2026;34(4):1318-1326. doi:10.1002/ksa.70037
BibTeX
@article{Montagna2026,
author = {Montagna, Alice and Marescalchi, Marina and Cinelli, Virginia and Sangaletti, Rudy and Andriollo, Luca and Benazzo, Francesco and Rossi, Stefano Marco Paolo},
title = {A novel knee implant for total knee arthroplasty meets expectations at 10 years. First long‐term follow‐up report of clinical outcomes and survivorship},
journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
year = {2026},
volume = {34},
number = {4},
pages = {1318--1326},
doi = {10.1002/ksa.70037}
}