Image‐based robotics enhance precision and efficiency in lateral UKA: A comparative study of 135 UKAs
Knee Surgery, Sports Traumatology, Arthroscopy · 2026
DOI: 10.1002/ksa.70012 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, 2026. Verfasst von Clément Favroul und 4 weiteren Autor:innen. Beteiligte Einrichtung: Orthopaedics Surgery and Sports Medicine Department, FIFA Medical Center of Excellence, Croix-Rousse Hospital, Hospices Civils de Lyon, Lyon North Un…. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Arthroplasty, Replacement, Knee, Osteoarthritis, Knee. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Robotic‐assisted lateral unicompartmental knee arthroplasty (UKA) remains technically demanding due to the complex biomechanics of the lateral compartment. Image‐based (IBRA) and imageless (ILRA) robotic systems have both demonstrated superior accuracy compared to conventional mechanical instrumentation, but have not yet been directly compared in lateral UKA. This study aimed to evaluate their respective accuracy and surgical efficiency.
Methods
This retrospective study included 135 patients who underwent lateral UKA using either IBRA or ILRA systems. Post‐operative radiographic outcomes included hip–knee–ankle (HKA) alignment, posterior tibial slope (PTS) and joint line (JL) restoration. Surgical time was assessed as a secondary outcome. Target zones were HKA 180°–185°, PTS 2°–8° and JL ± 2 mm.
Results
IBRA showed higher rates of HKA inliers (94.9% vs. 78.9%, p = 0.011) and JL restoration (76.2% vs. 31.6%, p < 0.001) compared to ILRA. PTS accuracy was similar between groups ( p = 0.30). Operative time was significantly shorter with IBRA (61.6 ± 13.5 vs. 81.9 ± 26.1 min, p < 0.001).
Conclusion
IBRA outperformed ILRA in terms of alignment accuracy and surgical duration. These findings support the added value of IBRA systems in lateral UKA. Level of Evidence Level IV, retrospective case series study.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Orthopaedics Surgery and Sports Medicine Department, FIFA Medical Center of Excellence, Croix-Rousse Hospital, Hospices Civils de Lyon, Lyon North University Hospital, Lyon, France.
- Univ Lyon, Claude Bernard Lyon 1 University, IFSTTAR, LBMC UMR_T9406, Lyon, France.
- Department of Orthopaedic Surgery and Traumatology, Faculty of Medecine, Minia University, Minia, Egypt.
- LIBM-EA 7424, Interuniversity Laboratory of Biology of Mobility, Claude Bernard Lyon 1 University, Lyon, France.
ORCID-Profile: Clément Favroul
Zitieren
APA
Favroul, C., Batailler, C., Abdelatif, E. A., Servien, E., & Lustig, S. (2026). Image‐based robotics enhance precision and efficiency in lateral UKA: A comparative study of 135 UKAs. Knee Surgery, Sports Traumatology, Arthroscopy. https://doi.org/10.1002/ksa.70012
Vancouver
Favroul C, Batailler C, Abdelatif EA, Servien E, Lustig S. Image‐based robotics enhance precision and efficiency in lateral UKA: A comparative study of 135 UKAs. Knee Surgery, Sports Traumatology, Arthroscopy. 2026. doi:10.1002/ksa.70012
BibTeX
@article{Favroul2026,
author = {Favroul, Clément and Batailler, Cécile and Abdelatif, Elsayed Ahmed and Servien, Elvire and Lustig, Sébastien},
title = {Image‐based robotics enhance precision and efficiency in lateral UKA: A comparative study of 135 UKAs},
journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
year = {2026},
doi = {10.1002/ksa.70012}
}
EUVIMED