EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieOpen Access · CC BY 4.0

Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step‐up and lunge motions

Lennard A Koster, Bart L Kaptein, Enrike H M J van der Linden-van der Zwaag, Rob G H H Nelissen

Knee Surgery, Sports Traumatology, Arthroscopy · 2024

DOI: 10.1002/ksa.12125 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, 2024. Verfasst von Lennard A Koster und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics, Biomechanics and Imaging Group, Leiden University Medical Center, Leiden, The Netherlands. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Knee Prosthesis, Prosthesis Design, Range of Motion, Articular. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Zimmer Biomet.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This clinical fluoroscopy study investigated knee kinematics of two different cemented fixed‐bearing, posterior‐stabilised (PS) total knee arthroplasty (TKA) designs: an asymmetric tibial component including an asymmetric insert designed to optimise personalised balance and fit and its precursor symmetrical design with symmetric insert.

Methods

A consecutive series of patients (16 TKAs from each treatment group) participating in a randomised controlled trial comparing TKA migration was included. The exclusion criterion was the use of walking aids. Flat‐panel fluoroscopic recordings of step‐up and lunge motions were acquired 1‐year postoperatively. Medial and lateral contact points (CPs) were determined to calculate CP displacement, femoral axial rotation and pivot position. Using linear mixed‐effects modelling techniques, kinematics between TKA designs were compared.

Results

During knee extension between 20° flexion and full extension, the CPs moved anteriorly combined with a small internal femoral rotation (a screw‐home mechanism). Whereas CP movement was reversed: femoral rollback, external femoral rotation while flexing the knee between full extension and 20° knee flexion, At larger flexion angles, femoral axial rotation (FAR) occurred around a lateral pivot point both during step‐up and lunge. The symmetric design had a 2.3° larger range of FAR compared to the asymmetric design during lunge ( p = 0.02). All other kinematics were comparable.

Conclusion

Despite the differences in design, this study showed that the asymmetric and symmetric PS TKA designs had mostly comparable knee kinematics during step‐up and lunge motions. It is therefore expected that the functionality of the successor TKA design is similar to that of its precursor design. Level of Evidence Level II, prospective comparative study.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeKnee ProsthesisProsthesis DesignRange of Motion, ArticularTibiaHumansFluoroscopyBiomechanical PhenomenaFemaleMaleAgedMiddle AgedKnee JointRotationOsteoarthritis, Knee

Stichwörter der Autor:innen

NexGenPersonafluoroscopic analysisknee kinematicsposterior‐stabilised TKAtotal knee arthroplasty

Autor:innen und Institutionen

  • Department of Orthopaedics, Biomechanics and Imaging Group, Leiden University Medical Center, Leiden, The Netherlands.

Förderung

  • Zimmer Biomet

Zitieren

APA

Koster, L. A., Kaptein, B. L., van der Linden-van der Zwaag, E. H. M. J., & Nelissen, R. G. H. H. (2024). Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step‐up and lunge motions. Knee Surgery, Sports Traumatology, Arthroscopy. https://doi.org/10.1002/ksa.12125

Vancouver

Koster LA, Kaptein BL, van der Linden-van der Zwaag EHMJ, Nelissen RGHH. Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step‐up and lunge motions. Knee Surgery, Sports Traumatology, Arthroscopy. 2024. doi:10.1002/ksa.12125

BibTeX
@article{Koster2024,
  author = {Koster, Lennard A and Kaptein, Bart L and van der Linden-van der Zwaag, Enrike H M J and Nelissen, Rob G H H},
  title = {Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step‐up and lunge motions},
  journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
  year = {2024},
  doi = {10.1002/ksa.12125}
}