In Vitro and In Vivo Evaluation of Ultrasound‐Triggered Release From Novel Spinal Device
Journal of Ultrasound in Medicine · 2023-10 · 42 (10) · 2357-2368
DOI: 10.1002/jum.16263 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Ultrasound in Medicine, online erschienen am 30.05.2023. Verfasst von Lauren J Delaney und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Vancomycin, Polyesters. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIAMS NIH HHS, National Institute of Arthritis and Musculoskeletal and Skin Diseases.
Abstract (Originalsprache)
Objectives
Bacterial infection following spinal fusion is a major clinical concern with up to 20% incidence. An ultrasound‐triggered bulk‐release system to combat postsurgical bacterial survival was designed and evaluated.
Methods
Polylactic acid (PLA) clips were loaded with vancomycin (VAN) and microbubbles (Sonazoid, GE HealthCare) in vitro. Stability was determined over 14 days. VAN‐loaded clips were submerged in water and insonated using a Logiq E10 scanner (GE HealthCare) with a curvilinear C6 probe. Doppler‐induced VAN release was quantified using spectrophotometry. For in vivo testing, clips were loaded with methylene blue (MeB) solution and Sonazoid. These clips were implanted into a rabbit along the spine at L2 and L5, as well as a pig at L1 and L3, then insonated in Doppler mode using the C6 probe.
Results
Sonazoid microbubbles were better preserved when incubated in VAN compared with distilled water at 4°C, 25°C, and 37°C incubation temperatures ( P = .0131). Contrast enhancement was observed from both solutions when incubated at 4°C storage conditions. Insonated clips achieved average cumulative VAN release of 101.8 ± 2.8% (81.4 ± 2.8 mg) after 72 hours. Uninsonated clips had only 0.3 ± 0.1% (0.3 ± 0.1 mg) average cumulative VAN release ( P < .0001). Clips retrieved from the rabbit did not rupture with insonation nor produce MeB staining of surrounding tissues. In the pig, the PLA film was visibly ruptured and MeB tissue was observed following insonation, whereas the uninsonated clip was intact.
Conclusion
These results demonstrate ultrasound‐triggered release of an encapsulated prophylactic solution and provide an important proof‐of‐concept for continuing large animal evaluations for translational merit.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
- Department of Orthopaedic Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
- Department of Clinical Studies, New Bolton Center, University of Pennsylvania School of Veterinary Medicine, Philadelphia, Pennsylvania, USA.
- School of Biomedical Engineering, Science & Health Systems, Drexel University, Philadelphia, Pennsylvania, USA.
- Exponent, Inc., Philadelphia, Pennsylvania, USA.
ORCID-Profile: Lauren J Delaney, Rachel Hilliard, Flemming Forsberg
Förderung
- NIAMS NIH HHS
- National Institute of Arthritis and Musculoskeletal and Skin Diseases
Zitieren
APA
Delaney, L. J., Isguven, S., Hilliard, R., Lacerda, Q., Oeffinger, B. E., Machado, P., Schaer, T. P., Hickok, N. J., Kurtz, S. M., & Forsberg, F. (2023). In Vitro and In Vivo Evaluation of Ultrasound‐Triggered Release From Novel Spinal Device. Journal of Ultrasound in Medicine, 42(10), 2357-2368. https://doi.org/10.1002/jum.16263
Vancouver
Delaney LJ, Isguven S, Hilliard R, Lacerda Q, Oeffinger BE, Machado P, et al. In Vitro and In Vivo Evaluation of Ultrasound‐Triggered Release From Novel Spinal Device. Journal of Ultrasound in Medicine. 2023;42(10):2357-2368. doi:10.1002/jum.16263
BibTeX
@article{Delaney2023,
author = {Delaney, Lauren J and Isguven, Selin and Hilliard, Rachel and Lacerda, Quezia and Oeffinger, Brian E and Machado, Priscilla and Schaer, Thomas P and Hickok, Noreen J and Kurtz, Steven M and Forsberg, Flemming},
title = {In Vitro and In Vivo Evaluation of Ultrasound‐Triggered Release From Novel Spinal Device},
journal = {Journal of Ultrasound in Medicine},
year = {2023},
volume = {42},
number = {10},
pages = {2357--2368},
doi = {10.1002/jum.16263}
}
EUVIMED