EUVIMEDEUVIMED entdecken
Crossref

Analytical quality by design approach for the development of high‐performance liquid chromatography method for simultaneous analysis of metformin and sitagliptin in the presence of major degradation products

Rajesh Pawar, Pothuraju Naresh iD, Bhadrayya Kalva, M Raj Kumar, V Ravichandiran, P Ramalingam

Journal of Separation Science · 2024-03 · 47 (5)

DOI: 10.1002/jssc.202300605 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Separation Science, online erschienen am 11.03.2024. Verfasst von Rajesh Pawar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmaceutical Analysis, National Institute of Pharmaceutical Education and Research (NIPER) Hajipur, Bihar, India. Hauptthemen laut MEDLINE: Sitagliptin Phosphate, Metformin, Phosphoric Acids. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 25 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

An analytical quality by design‐based high‐performance liquid chromatography method for determining metformin (MET) and sitagliptin (SIT) in stress‐degraded samples was developed and validated. The analytical target profile and risk assessment‐driven critical method variables, for example, pH, % aqueous, and buffer concentration, were studied for their effect on method responses of retention time and resolution using a central composite design. The correlation regression coefficient was more than 0.8, and variables interaction was significant on method responses with curvature effect. The method operable design region afforded an aqueous range of 55%–70% and an ortho ‐phosphoric acid buffer of 0.1% with a pH of 3.0–4.0 as a robust region for the suitable method performance characteristics. The separation of MET and SIT from their degradants ( m/z 85.0509; m/z 193.0694) on the C 8 column was achieved using a mobile phase consisting of 0.1% ortho ‐phosphoric acid and methanol (60:40% v/v; pH 3.0). The optimized method eluted MET and SIT at 4.3 ± 0.2 and 7.1 ± 0.2 min, respectively, with acceptable specificity and resolution. The linearity ranges of 25–250 μg/mL (r 2 : 0.9982) and 5–50 μg/mL (r 2 : 0.9989) was established for MET and SIT, respectively. The % recovery (98.81%–102.17%), precision (0.55%–1.65%), and robustness study for method variables were acceptable.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Sitagliptin PhosphateMetforminPhosphoric AcidsChromatography, High Pressure LiquidMethanol

Stichwörter der Autor:innen

critical impuritiesmetforminquality by designsitagliptinstability

Substanzen

Sitagliptin Phosphatephosphoric acidMetforminMethanolPhosphoric Acids

Autor:innen und Institutionen

  • Department of Pharmaceutical Analysis, National Institute of Pharmaceutical Education and Research (NIPER) Hajipur, Bihar, India.
  • Swaroop Tech Services Private Ltd, Hyderabad, India.
  • Department of Pharmaceutics, National Institute of Pharmaceutical Education and Research (NIPER) Hajipur, Bihar, India.

ORCID-Profile: Pothuraju Naresh

Zitieren

APA

Pawar, R., Naresh, P., Kalva, B., Kumar, M. R., Ravichandiran, V., & Ramalingam, P. (2024). Analytical quality by design approach for the development of high‐performance liquid chromatography method for simultaneous analysis of metformin and sitagliptin in the presence of major degradation products. Journal of Separation Science, 47(5). https://doi.org/10.1002/jssc.202300605

Vancouver

Pawar R, Naresh P, Kalva B, Kumar MR, Ravichandiran V, Ramalingam P. Analytical quality by design approach for the development of high‐performance liquid chromatography method for simultaneous analysis of metformin and sitagliptin in the presence of major degradation products. Journal of Separation Science. 2024;47(5). doi:10.1002/jssc.202300605

BibTeX
@article{Pawar2024,
  author = {Pawar, Rajesh and Naresh, Pothuraju and Kalva, Bhadrayya and Kumar, M Raj and Ravichandiran, V and Ramalingam, P},
  title = {Analytical quality by design approach for the development of high‐performance liquid chromatography method for simultaneous analysis of metformin and sitagliptin in the presence of major degradation products},
  journal = {Journal of Separation Science},
  year = {2024},
  volume = {47},
  number = {5},
  doi = {10.1002/jssc.202300605}
}
Kostenlos starten →