Improved Wear Performance of Highly Cross‐Linked Versus Conventional Polyethylene in Dual Mobility Cups
Journal of Orthopaedic Research · 2026-11 · 44 (11)
DOI: 10.1002/jor.70313 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Orthopaedic Research, online erschienen am 17.11.2026. Verfasst von Cristina Cocor und 4 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Mobilisation
Abstract (Originalsprache)
The purpose of the present study was to compare wear between vitamin E infused highly cross‐linked polyethylene (vit E XLPE) versus ultra‐high molecular weight polyethylene (UHMWPE) dual‐mobility (DM) mobile components for total hip arthroplasty (THA), in worst‐case conditions. Four in vitro tests were performed: (1) global wear test with pristine mobile components, (2) global wear test with accelerated ageing mobile components, (3) outer bearing only wear test with pristine mobile components, and (4) outer bearing only wear test with pristine mobile components and third‐body particles. For each of these four tests, three vit E XLPE and three UHMWPE DM mobile components were tested, with the largest available DM cup size. Wear was calculated using the gravimetric method, adjusting for fluid uptake. Total wear at 5 million cycles was significantly lower for vit E XLPE compared to UHMWPE for all four tests: 79.3 ± 9.6 mg vs. 138.2 ± 7.2 mg ( p = 0.003) for global wear with pristine mobile components, 54.9 ± 3.0 mg vs. 592.1 ± 79.6 mg ( p = 0.007) for global wear with accelerated ageing mobile components, 59.8 ± 10.2 mg vs. 231.6 ± 15.0 mg ( p < 0.001) for outer bearing wear with pristine mobile components, and 75.4 ± 17.5 mg vs. 370.5 ± 46.1 mg ( p = 0.009) for outer bearing wear with pristine mobile components and third‐body particles. In conclusion, vit E XLPE DM mobile components had significantly reduced wear compared to UHMWPE DM mobile components. Difference in wear was greatest for accelerated ageing and third‐body particles tests, which may be most relevant for young and active patients, as well as for revision THA cases. Clinical testing of DM systems is necessary to confirm these findings in vivo.
Zitieren
APA
Cocor, C., Sellier, A., Imbert, C., Ramos‐Pascual, S., & Wegrzyn, J. (2026). Improved Wear Performance of Highly Cross‐Linked Versus Conventional Polyethylene in Dual Mobility Cups. Journal of Orthopaedic Research, 44(11). https://doi.org/10.1002/jor.70313
Vancouver
Cocor C, Sellier A, Imbert C, Ramos‐Pascual S, Wegrzyn J. Improved Wear Performance of Highly Cross‐Linked Versus Conventional Polyethylene in Dual Mobility Cups. Journal of Orthopaedic Research. 2026;44(11). doi:10.1002/jor.70313
BibTeX
@article{Cocor2026,
author = {Cocor, Cristina and Sellier, Agathe and Imbert, Cyril and Ramos‐Pascual, Sonia and Wegrzyn, Julien},
title = {Improved Wear Performance of Highly Cross‐Linked Versus Conventional Polyethylene in Dual Mobility Cups},
journal = {Journal of Orthopaedic Research},
year = {2026},
volume = {44},
number = {11},
doi = {10.1002/jor.70313}
}
EUVIMED