Mechanical Evaluation of Posterior Wall Acetabular Facture Fixation: Patient‐Specific Implants Versus Conventional Implants
Journal of Orthopaedic Research · 2026
DOI: 10.1002/jor.70062 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Orthopaedic Research, 2026. Verfasst von Miriam G E Oldhoff und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Trauma Surgery, University Medical Centre Groningen, University of Groningen, Groningen, The Netherlands. Hauptthemen laut MEDLINE: Acetabulum, Fractures, Bone, Fracture Fixation, Internal. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: This study was supported, within the Just Transition Fund program IMPACT-implants, co-financed by the European Union, with co-financing from the Ministry of Ec….
Abstract (Originalsprache)
The aim of the study was to design a standardized mechanical test setup and a corresponding finite element analysis to assess the stability and strength of both patient‐specific and conventional implants for posterior wall acetabular fractures. Ten synthetic hemi‐pelves with posterior wall fractures were biomechanically tested with two types of implants: a patient‐specific implant (PSI) and a seven‐hole plate conventional implant. 3D‐printed guides ensured reproducibility. The models were tested using an Instron machine. The protocol involved 10,000 cyclic load cycles with static tests at 3200 N before and after to simulate early postoperative weightbearing conditions. Construct stiffness, stiffness over cyclic loading and fracture gapping were measured and compared. A finite element analysis was created with similar conditions to investigate stresses within the synthetic bone and fixation materials. The mechanical tests showed comparable stiffness for PSI (1.75 kN/mm) and the conventional implant (1.71 kN/mm, p = 0.47). Stability over 10,000 cycles was similar, and fracture gapping remained minimal (0.0–0.8 mm) without significant differences. No failure or plastic deformation occurred under 3200 N loading. Finite element analysis confirmed that von Mises stresses remained below the yield stress. This study introduces a reproducible workflow for biomechanical testing of acetabular fractures using synthetic bone models and 3D‐printed guides. It serves as a step‐by‐step guideline and standard reference for pelvic biomechanical testing. Both patient‐specific and conventional implants, using a seven‐hole plate construct with one or two screws through the plate in the fracture fragment, provide stable fixation for large posterior wall fragments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Trauma Surgery, University Medical Centre Groningen, University of Groningen, Groningen, The Netherlands.
- 3D Lab, University Medical Centre Groningen, University of Groningen, Groningen, The Netherlands.
- Department of Oral and Maxillofacial Surgery, University Medical Centre Groningen, University of Groningen, Groningen, The Netherlands.
- Pelvic and Acetabular Reconstruction Unit, Southmead Hospital, North Bristol NHS Trust, Bristol, UK.
- Department of Surgery, Division of Vascular Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
ORCID-Profile: Miriam G E Oldhoff, Anne M L Meesters, Joep Kraeima, Kaj Ten Duis, Jean-Paul P M de Vries, Bram B J Merema, Frank F A IJpma
Förderung
- This study was supported, within the Just Transition Fund program IMPACT-implants, co-financed by the European Union, with co-financing from the Ministry of Ec…
Zitieren
APA
Oldhoff, M. G. E., Meesters, A. M. L., Kraeima, J., Acharya, M. R., Ten Duis, K., de Vries, J. P. M., Merema, B. B. J., & IJpma, F. F. A. (2026). Mechanical Evaluation of Posterior Wall Acetabular Facture Fixation: Patient‐Specific Implants Versus Conventional Implants. Journal of Orthopaedic Research. https://doi.org/10.1002/jor.70062
Vancouver
Oldhoff MGE, Meesters AML, Kraeima J, Acharya MR, Ten Duis K, de Vries JPM, et al. Mechanical Evaluation of Posterior Wall Acetabular Facture Fixation: Patient‐Specific Implants Versus Conventional Implants. Journal of Orthopaedic Research. 2026. doi:10.1002/jor.70062
BibTeX
@article{Oldhoff2026,
author = {Oldhoff, Miriam G E and Meesters, Anne M L and Kraeima, Joep and Acharya, Mehool R and Ten Duis, Kaj and de Vries, Jean-Paul P M and Merema, Bram B J and IJpma, Frank F A},
title = {Mechanical Evaluation of Posterior Wall Acetabular Facture Fixation: Patient‐Specific Implants Versus Conventional Implants},
journal = {Journal of Orthopaedic Research},
year = {2026},
doi = {10.1002/jor.70062}
}
EUVIMED