Effects of Plate‐Bone Contact on Bridge Plate Fracture Fixation Biomechanics: Computational, Experimental, and Analytical Modeling
Journal of Orthopaedic Research · 2025-09 · 43 (9) · 1606-1618
DOI: 10.1002/jor.26114 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Orthopaedic Research, online erschienen am 08.06.2025. Verfasst von Zachary A Koroneos und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Rehabilitation, Penn State College of Medicine, Hershey, Pennsylvania, USA. Hauptthemen laut MEDLINE: Bone Plates, Fracture Fixation, Internal. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 67 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NIBIB NIH HHS, This study was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under award number R01EB029207..
Abstract (Originalsprache)
Bridge plating is commonly used for internal fixation of comminuted fractures. The inner working length between screws has been established as a key parameter controlling postoperative biomechanical stability. However, plate‐bone contact may affect these biomechanics in complex ways, and the offset between the plate and bone is variable across surgeries. The objective of this study was to examine the effects of construct and loading parameters on interfragmentary motion and maximum plate stress of bridge plating constructs. Finite element models were developed with variations in inner working length, plate‐bone offset, fracture gap size, and loading type and magnitudes. Experiments with synthetic bones were conducted in parallel to support model credibility. Analytical models were also developed based on beam bending and torsion of the plate, assuming rigidity outside the inner working length. Finite element and experimental results of axial and torsional loading scenarios without plate‐bone contact confirmed linear relationships between inner working length and interfragmentary motion. Analytical predictions of interfragmentary motion showed very good agreement with the finite element simulations in these scenarios. Conversely, in cases with plate‐bone contact, a shorter effective working length was formed, and results were dependent on additional variables such as fracture gap. The study shows how the mechanics of bridge plating can be understood and predicted based on beam theory up to the point of plate‐bone contact, and how interfragmentary motions and maximum plate stresses are influenced by the interaction of surgical variables in the presence of plate‐bone contact.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics and Rehabilitation, Penn State College of Medicine, Hershey, Pennsylvania, USA.
- Department of Mechanical Engineering, The Pennsylvania State University, University Park, Pennsylvania, USA.
ORCID-Profile: Gregory S Lewis
Förderung
- NIBIB NIH HHS
- This study was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under award number R01EB029207.
Zitieren
APA
Koroneos, Z. A., Wee, H., Reid, J. S., & Lewis, G. S. (2025). Effects of Plate‐Bone Contact on Bridge Plate Fracture Fixation Biomechanics: Computational, Experimental, and Analytical Modeling. Journal of Orthopaedic Research, 43(9), 1606-1618. https://doi.org/10.1002/jor.26114
Vancouver
Koroneos ZA, Wee H, Reid JS, Lewis GS. Effects of Plate‐Bone Contact on Bridge Plate Fracture Fixation Biomechanics: Computational, Experimental, and Analytical Modeling. Journal of Orthopaedic Research. 2025;43(9):1606-1618. doi:10.1002/jor.26114
BibTeX
@article{Koroneos2025,
author = {Koroneos, Zachary A and Wee, Hwabok and Reid, J Spence and Lewis, Gregory S},
title = {Effects of Plate‐Bone Contact on Bridge Plate Fracture Fixation Biomechanics: Computational, Experimental, and Analytical Modeling},
journal = {Journal of Orthopaedic Research},
year = {2025},
volume = {43},
number = {9},
pages = {1606--1618},
doi = {10.1002/jor.26114}
}