Influence of sidewall retention and interference fit in total ankle replacement on implant‐bone micromotion: A finite element study
Journal of Orthopaedic Research · 2024
DOI: 10.1002/jor.25796 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Orthopaedic Research, 2024. Verfasst von Joshua E Johnson und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics and Rehabilitation, The University of Iowa, Iowa City, Iowa, USA. Hauptthemen laut MEDLINE: Finite Element Analysis, Arthroplasty, Replacement, Ankle. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Paragon 28.
Abstract (Originalsprache)
The success of uncemented total ankle replacement (TAR) is linked to initial stability because bony ingrowth depends upon limited early micromotion. Tibial implant design fixation features resist micromotion aided by bony sidewall retention and interference fit. Our goal was to investigate factors influencing implant‐bone micromotion in TAR. Two TAR tibial components were virtually inserted into CT‐derived computer models of two distal tibias from patients with end‐stage ankle arthritis. Density‐based inhomogeneous material assignment was used to model bone compaction during press‐fit. Finite element analysis (FEA) was used to simulate three fixation cases: (1) no sidewalls + line‐to‐line fit, (2) sidewalls + line‐to‐line fit, and (3) sidewalls + 50, 100, or 200 µm interference fit. Kinetic profiles from the stance phase of gait were simulated and micromotions computed from FEA output. Without sidewalls or interference fit, micromotions were largest in early and late stance, with largest micromotions (averaging ~150–250 µm) observed near heel strike. Micromotions decreased 39%–62% when sidewalls were retained. When interference fit was also modeled, micromotions decreased another 37%–61% to ~10 µm. Micromotion differences between patients persisted with sidewall retention but largely disappeared with interference fit. This study presents new insights into the effects of TAR fixation features on implant‐bone micromotion. Stability appeared to be influenced by surrounding bone quality, but this influence was greatly diminished when interference fit was introduced. More complete understanding of TAR implant features and performance is needed, but our results show the importance of bone quality and interference fit in the stability of uncemented TAR.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedics and Rehabilitation, The University of Iowa, Iowa City, Iowa, USA.
- Department of Biomedical Engineering, The University of Iowa, Iowa City, Iowa, USA.
ORCID-Profile: Joshua E Johnson, Donald D Anderson
Förderung
- Paragon 28
Zitieren
APA
Johnson, J. E., Clarke, G. A., de Cesar Netto, C., & Anderson, D. D. (2024). Influence of sidewall retention and interference fit in total ankle replacement on implant‐bone micromotion: A finite element study. Journal of Orthopaedic Research. https://doi.org/10.1002/jor.25796
Vancouver
Johnson JE, Clarke GA, de Cesar Netto C, Anderson DD. Influence of sidewall retention and interference fit in total ankle replacement on implant‐bone micromotion: A finite element study. Journal of Orthopaedic Research. 2024. doi:10.1002/jor.25796
BibTeX
@article{Johnson2024,
author = {Johnson, Joshua E and Clarke, Gabriel A and de Cesar Netto, Cesar and Anderson, Donald D},
title = {Influence of sidewall retention and interference fit in total ankle replacement on implant‐bone micromotion: A finite element study},
journal = {Journal of Orthopaedic Research},
year = {2024},
doi = {10.1002/jor.25796}
}