Clinical validation of novel dried blood spot based collecting device using serum separation for measuring SARS‐CoV‐2 antibodies
Journal of Medical Virology · 2023
DOI: 10.1002/jmv.28765 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Virology, 2023. Verfasst von Sophie Jolien Schuurmans Stekhoven und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Internal Medicine, Spaarne Gasthuis, Haarlem, The Netherlands. Hauptthemen laut MEDLINE: COVID-19. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 13 Literaturangaben.
Abstract (Originalsprache)
Accurate surveillance of coronavirus disease 2019 (COVID‐19) incidence includes large‐scale antibody testing of the population. Current testing methods require collection of venous blood samples by a healthcare worker, or dried blood spot (DBS) collection using finger prick, however this might have some logistic and processing limitations. We investigated the performance of the Ser‐Col device for detecting severe acute respiratory syndrome coronavirus‐2 (SARS‐CoV‐2) antibodies using a finger prick: DBS‐like collection system that includes a lateral flow paper for serum separation and allows for automated large scale analysis. For this prospective study, adult patients with moderate to severe COVID‐19 were included 6 weeks post‐symptom onset. Healthy, adult volunteers were included as a negative control group. Venous blood and capillary blood using the Ser‐Col device were collected and the Wantai SARS‐CoV‐2 total antibody ELISA was performed on all samples. We included 50 subjects in the study population and 49 in the control group. Results obtained with venous blood versus Ser‐Col capillary blood showed 100% sensitivity (95% CI: 0.93−1.00) and 100% specificity (95% CI: 0.93−1.00). Our study shows the feasibility of SARS‐CoV‐2 total antibody screening using a standardized DBS technique with semiautomated processing for large scale analysis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Internal Medicine, Spaarne Gasthuis, Haarlem, The Netherlands.
- Regional Public Health Laboratory Kennemerland, Haarlem, The Netherlands.
- Department of Pulmonology, Spaarne Gasthuis, Haarlem, The Netherlands.
- Department of Paediatrics, Spaarne Gasthuis, Haarlem, The Netherlands.
ORCID-Profile: Sophie Jolien Schuurmans Stekhoven
Zitieren
APA
Schuurmans Stekhoven, S. J., Winkel, K. G. T., Souverein, D., Sondermeijer, B. M., van Houten, M. A., Euser, S. M., & van Lelyveld, S. F. L. (2023). Clinical validation of novel dried blood spot based collecting device using serum separation for measuring SARS‐CoV‐2 antibodies. Journal of Medical Virology. https://doi.org/10.1002/jmv.28765
Vancouver
Schuurmans Stekhoven SJ, Winkel KGT, Souverein D, Sondermeijer BM, van Houten MA, Euser SM, et al. Clinical validation of novel dried blood spot based collecting device using serum separation for measuring SARS‐CoV‐2 antibodies. Journal of Medical Virology. 2023. doi:10.1002/jmv.28765
BibTeX
@article{SchuurmansStekhoven2023,
author = {Schuurmans Stekhoven, Sophie Jolien and Winkel, Kirsten Gerdien Te and Souverein, Dennis and Sondermeijer, Brigitte Marlene and van Houten, Marianne A and Euser, Sjoerd M and van Lelyveld, Steven Ferdinand Lodewijk},
title = {Clinical validation of novel dried blood spot based collecting device using serum separation for measuring SARS‐CoV‐2 antibodies},
journal = {Journal of Medical Virology},
year = {2023},
doi = {10.1002/jmv.28765}
}
EUVIMED