EUVIMEDEUVIMED entdecken
Validierungsstudie

The Combined Quantification of the 15 Human Bile Acids and Their Precursor 7α‐Hydroxy‐4‐Cholesten‐3‐One in Human Serum: LC‐MS/MS Method Development and Validation

Zuweina Al-Hadidi, Sathish Pandian, Mohammed Al-Tobi, Fahad Al-Zadjali, Amani Al-Khaifi iD

Journal of Mass Spectrometry · 2026-06 · 61 (6)

DOI: 10.1002/jms.70064 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Mass Spectrometry, online erschienen am 13.05.2026. Verfasst von Zuweina Al-Hadidi und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biochemistry, College of Medicine and Health Sciences, Sultan Qaboos University, Muscat, Oman. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Liquid Chromatography-Mass Spectrometry, Bile Acids and Salts, Cholestenones. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Ministry of Higher Education, Research and Innovation, Ministère de l'Enseignement supérieur, de la Recherche et de l'Innovation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Bile acids (BAs) have been considered potential biomarkers for various metabolic diseases, which prompted the need for accurate and sensitive quantification methods. This study aimed to develop a sensitive, validated liquid chromatography–tandem mass spectrometry (LC‐MS/MS) method for quantifying primary, secondary, free, and conjugated BAs, along with their precursor, 7α‐hydroxy‐4‐cholesten‐3‐one (C4), in human serum. The analytical approach involved a rapid protein precipitation‐based extraction technique in the presence of deuterated internal standards. Mass spectrometry detection employed electrospray ionization in negative and positive ion modes for BAs and C4, respectively. BA levels obtained by LC‐MS/MS were validated by comparing them to total BAs measured by a sensitive enzymatic assay. The chromatographic cycle time for the method was 21 min per injection, employing multiple reaction monitoring. All BAs were separated with negligible carryover, matrix effect, and good recovery. Good linearity was observed in the concentration range of (0.0–10.0) μM for BAs and (0.0–0.4) μM for C4. The detection limit for BAs was between 0.0002 and 0.0138 μmol/L, and the sensitivity was 0.00156–0.0125 μmol/L. Acceptable precision results were obtained on intraday and interday analyses for BAs and C4. The LC‐MS/MS quantification of total BAs closely matched levels from the colorimetric assay. This study presents a straightforward LC‐MS/MS method for simultaneously quantifying BAs and their precursor, C4, in human serum. The low sample volume, rapid extraction, validated results, and detailed protocol make the established method reproducible and suitable for use.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tandem Mass SpectrometryLiquid Chromatography-Mass SpectrometryBile Acids and SaltsCholestenonesHumansReproducibility of ResultsLimit of DetectionLinear Models

Stichwörter der Autor:innen

CYP7A1 biomarkerhuman bile acidsliquid chromatography–tandem mass spectrometry

Substanzen

Bile Acids and Salts7 alpha-hydroxy-4-cholesten-3-oneCholestenones

Autor:innen und Institutionen

  • Department of Biochemistry, College of Medicine and Health Sciences, Sultan Qaboos University, Muscat, Oman.
  • Central Analytical and Applied Research Unit, College of Science, Sultan Qaboos University, Muscat, Oman.
  • Oman College of Health Sciences, Muscat, Oman.

ORCID-Profile: Amani Al-Khaifi

Förderung

  • Ministry of Higher Education, Research and Innovation
  • Ministère de l'Enseignement supérieur, de la Recherche et de l'Innovation

Zitieren

APA

Al-Hadidi, Z., Pandian, S., Al-Tobi, M., Al-Zadjali, F., & Al-Khaifi, A. (2026). The Combined Quantification of the 15 Human Bile Acids and Their Precursor 7α‐Hydroxy‐4‐Cholesten‐3‐One in Human Serum: LC‐MS/MS Method Development and Validation. Journal of Mass Spectrometry, 61(6). https://doi.org/10.1002/jms.70064

Vancouver

Al-Hadidi Z, Pandian S, Al-Tobi M, Al-Zadjali F, Al-Khaifi A. The Combined Quantification of the 15 Human Bile Acids and Their Precursor 7α‐Hydroxy‐4‐Cholesten‐3‐One in Human Serum: LC‐MS/MS Method Development and Validation. Journal of Mass Spectrometry. 2026;61(6). doi:10.1002/jms.70064

BibTeX
@article{AlHadidi2026,
  author = {Al-Hadidi, Zuweina and Pandian, Sathish and Al-Tobi, Mohammed and Al-Zadjali, Fahad and Al-Khaifi, Amani},
  title = {The Combined Quantification of the 15 Human Bile Acids and Their Precursor 7α‐Hydroxy‐4‐Cholesten‐3‐One in Human Serum: LC‐MS/MS Method Development and Validation},
  journal = {Journal of Mass Spectrometry},
  year = {2026},
  volume = {61},
  number = {6},
  doi = {10.1002/jms.70064}
}
Kostenlos starten →