EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Switch It Off! Carbon, Financial and Health Service Impacts of Switching Off a Computed Tomography Scanner: A Quality Improvement Study

Nicholas Marks iD, Mya Abigail Acosta, Kristen Pickles iD, Scott McAlister, Katy J L Bell iD, NSW Health Net Zero Clinical Leads Program

Journal of Medical Radiation Sciences · 2025-12 · 72 (4) · 430-438

DOI: 10.1002/jmrs.896 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Medical Radiation Sciences, online erschienen am 21.09.2025. Verfasst von Nicholas Marks und 5 weiteren Autor:innen. Beteiligte Einrichtung: NSW Health Net Zero Clinical Lead, Climate Risk and Net Zero Unit, Ministry of Health, St Leonards, New South Wales, Australia. Hauptthemen laut MEDLINE: Quality Improvement, Tomography Scanners, X-Ray Computed, Carbon, Tomography, X-Ray Computed. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NSW Health, National Health and Medical Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Medical imaging has been identified as a carbon hotspot in health care, and demand for imaging services is increasing. This study investigated switching off a surplus computerised tomography (CT) scanner when not clinically required as a possible simple and scalable intervention to reduce healthcare emissions.

Methods

This before‐after quality improvement study introduced a ‘Switch it off’ intervention where radiography staff switched off the power to a surplus CT scanner after hours (17:00–08:00) for 7 days (‘intervention period’: 07/7/2023–13/07/2023). Using a power data logger, power consumption (kilowatt hours, KWh) during the intervention period was compared to 7 days without the switch‐off practice (‘control period’: 24/07/2023–31/07/2023). Financial and carbon emission impacts were calculated based on energy consumption. All CT radiographers working in the department were invited to undertake a pre and post intervention survey. Differences in quantitative data pre‐ and postintervention were analysed using chi‐squared test for independent proportions. Free text survey responses were summarised into themes.

Results

Compared with energy use in the control period (433.96 kWh), there was a reduction in 139.79 kWh during the intervention period (294.17 kWh): 32% relative reduction. Extrapolation to 12 months found potential savings of 7280 kWh in energy use, $1381 to the hospital budget, and 5.5 T CO 2 e to the carbon budget. Of the 22 CT radiographers invited, 10 (45%) completed the survey, reporting no or trivial clinical impacts from switching off. The proportion of radiographers reporting switching off the scanner when not in use increased by 70% (95% CI: 39%, 100%; p = 0.002) from 10% (1/10) pre‐ to 80% (8/10) postintervention.

Conclusion

Identifying and switching off surplus CT scanners in low use times is a simple and scalable intervention that can achieve significant power, financial and carbon savings with little to no impact on clinical workflow.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality ImprovementTomography Scanners, X-Ray ComputedCarbonTomography, X-Ray ComputedHealth ServicesHumans

Stichwörter der Autor:innen

carbon footprintcomputerised tomographydiagnostic imaginggreenhouse gas emissionsnet zeropower consumptionradiography

Substanzen

Carbon

Autor:innen und Institutionen

  • NSW Health Net Zero Clinical Lead, Climate Risk and Net Zero Unit, Ministry of Health, St Leonards, New South Wales, Australia.
  • Calvary Mater Hospital, Hunter New England Imaging, NSW Health, Newcastle, New South Wales, Australia.
  • Wiser Healthcare Research Collaboration, School of Public Health, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
  • Healthy Environments and Lives (HEAL) Global Research Centre, Faculty of Health, Health Research Institute, University of Canberra, Canberra, Australian Capital Territory, Australia.
  • The Healthcare Carbon Lab, Department of Critical Care, University of Melbourne, Parkville, Victoria, Australia.

ORCID-Profile: Nicholas Marks, Kristen Pickles, Katy J L Bell

Förderung

  • NSW Health
  • National Health and Medical Research Council

Zitieren

APA

Marks, N., Acosta, M. A., Pickles, K., McAlister, S., Bell, K. J. L., & NSW Health Net Zero Clinical Leads Program (2025). Switch It Off! Carbon, Financial and Health Service Impacts of Switching Off a Computed Tomography Scanner: A Quality Improvement Study. Journal of Medical Radiation Sciences, 72(4), 430-438. https://doi.org/10.1002/jmrs.896

Vancouver

Marks N, Acosta MA, Pickles K, McAlister S, Bell KJL, NSW Health Net Zero Clinical Leads Program. Switch It Off! Carbon, Financial and Health Service Impacts of Switching Off a Computed Tomography Scanner: A Quality Improvement Study. Journal of Medical Radiation Sciences. 2025;72(4):430-438. doi:10.1002/jmrs.896

BibTeX
@article{Marks2025,
  author = {Marks, Nicholas and Acosta, Mya Abigail and Pickles, Kristen and McAlister, Scott and Bell, Katy J L and NSW Health Net Zero Clinical Leads Program, NSW Health Net Zero Clinical Leads Program},
  title = {Switch It Off! Carbon, Financial and Health Service Impacts of Switching Off a Computed Tomography Scanner: A Quality Improvement Study},
  journal = {Journal of Medical Radiation Sciences},
  year = {2025},
  volume = {72},
  number = {4},
  pages = {430--438},
  doi = {10.1002/jmrs.896}
}
Kostenlos starten →