Autonomous Velocity Encoding ( VENC ) Selection Improves Precision of Quantitative Flow Measurement
Journal of Magnetic Resonance Imaging · 2026
DOI: 10.1002/jmri.70443 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2026. Verfasst von Pierre Daudé und 9 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Imaging Technology, Cardiovascular Branch, Division of Intramural Research, National Heart, Lung & Blood Institute, NIH, Bethesda, Mary…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NHLBI NIH HHS, National Heart, Lung, and Blood Institute.
Abstract (Originalsprache)
Background
The optimal velocity encoding limit (VENC) in phase contrast MRI is subject‐specific because it depends on peak flow rate and the presence of flow jets. Currently, VENC is set manually with limited prior knowledge of the peak velocity. Setting the correct VENC might improve measurement precision or shorten scan times.
Purpose
To evaluate a workflow for autonomous selection of the optimal VENC without technologist interaction. Study Type Prospective and retrospective. Population The retrospective cohort included 254 scans from 113 patients (47 ± 14 years, 58 female) and 18 healthy volunteers (34 ± 15 years, 10 female). The prospective cohort included 5 patients with abnormal flow (47 ± 19 years, 2 female) and 10 healthy volunteers (28 ± 7 years, 6 female). Field Strength/Sequence 1.5 T, 0.55 T; Cartesian gradient echo phase‐contrast sequence. Assessment A workflow has been designed to autonomously estimate and prescribe the optimal VENC. A 30s free‐breathing calibration scan was followed by automatic estimation of the maximum velocity (Vmax) in < 6 s. The calculated optimal VENC was then automatically applied in the subsequent flow measurement without operator intervention. The target VENC:Vmax ratio was 1.1–1.25 according to consensus statements. Statistical Tests Shapiro–Wilk test, nonparametric two‐sided bootstrap with 90% confidence intervals, Wilcoxon signed‐rank test, paired t ‐test. Holm correction was applied to prespecified pairwise comparisons. p < 0.05 was considered significant.
Results
Retrospective analysis demonstrated sub‐optimal VENC setting in 96.5% of examinations with a VENC:Vmax ratio [2.03, 2.13] (two‐sided bootstrap 90% confidence interval). In the prospective cohort, autonomous inline VENC selection yielded a mean VENC:Vmax ratio of 1.19 ± 0.10, significantly lower than subject‐invariant VENC settings (VENC 150 :Vmax = 1.49 ± 0.31, VENC 200 :Vmax = 1.99 ± 0.41). Optimized VENC improved measurement precision, reducing velocity standard deviation by 41.9% ± 8.0%, and enabling shorter scan time to approximately 1/3 compared with default VENC 200 cm/s while maintaining equivalent velocity‐to‐noise performance. Data Conclusion Inline autonomous VENC selection improved flow‐measurement precision, simplified the acquisition workflow, and reduced scan time at 0.55 T. Evidence Level 2. Technical Efficacy 1.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratory of Imaging Technology, Cardiovascular Branch, Division of Intramural Research, National Heart, Lung & Blood Institute, NIH, Bethesda, Maryland, USA.
- Cardiovascular Branch, Division of Intramural Research, National Heart, Lung & Blood Institute, NIH, Bethesda, Maryland, USA.
- Siemens Medical Solutions USA Inc, Malvern, Pennsylvania, USA.
- Siemens Healthcare Ltd, Calgary, Alberta, Canada.
ORCID-Profile: Pierre Daudé, Christine Mancini, Dominique Franson, Sunggun Lee, Adrienne E Campbell-Washburn
Förderung
- NHLBI NIH HHS
- National Heart, Lung, and Blood Institute
Zitieren
APA
Daudé, P., Ramasawmy, R., Mancini, C., Franson, D., Lee, S., Tsakirellis, A., Bandettini, W. P., Javed, A., Chow, K., & Campbell-Washburn, A. E. (2026). Autonomous Velocity Encoding ( VENC ) Selection Improves Precision of Quantitative Flow Measurement. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70443
Vancouver
Daudé P, Ramasawmy R, Mancini C, Franson D, Lee S, Tsakirellis A, et al. Autonomous Velocity Encoding ( VENC ) Selection Improves Precision of Quantitative Flow Measurement. Journal of Magnetic Resonance Imaging. 2026. doi:10.1002/jmri.70443
BibTeX
@article{Daud2026,
author = {Daudé, Pierre and Ramasawmy, Rajiv and Mancini, Christine and Franson, Dominique and Lee, Sunggun and Tsakirellis, Anastasia and Bandettini, W Patricia and Javed, Ahsan and Chow, Kelvin and Campbell-Washburn, Adrienne E},
title = {Autonomous Velocity Encoding ( VENC ) Selection Improves Precision of Quantitative Flow Measurement},
journal = {Journal of Magnetic Resonance Imaging},
year = {2026},
doi = {10.1002/jmri.70443}
}
EUVIMED