Comparison Between Respiratory‐Triggered and Free‐Breathing Signal Acquisition Methods for Single‐Voxel Phosphorus Magnetic Resonance Spectroscopy of the Liver on a Clinical System
Journal of Magnetic Resonance Imaging · 2026-09 · 64 (3) · 702-711
DOI: 10.1002/jmri.70352 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Magnetic Resonance Imaging, online erschienen am 02.05.2026. Verfasst von Kumi Ozaki und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Hamamatsu University School of Medicine, Hamamatsu, Japan. Hauptthemen laut MEDLINE: Liver, Phosphorus, Respiratory-Gated Imaging Techniques. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The optimal protocol for clinical liver 31 P‐magnetic resonance spectroscopy (MRS) remains unclear. Single‐voxel 31 P‐MRS using image‐selected in vivo spectroscopy (ISIS) employs respiratory‐triggering (RT) or free‐breathing (FB) acquisition. RT provides robust data but prolongs scan duration; FB allows faster acquisition but may suffer from low signal‐to‐noise ratio (SNR). Yet, direct comparison using a regulatory‐approved coil in a clinical setting has not been reported.
Purpose
To compare 31 P‐MRS data stability and robustness between RT and FB. Study Type Prospective. Population 24 volunteers (19 male/5 female). Field Strength/Sequence 3 T MRI; single‐voxel 31 P‐MRS using ISIS (free induction decay‐based) with approved 31 P coil.
Methods
31 P‐MRS was performed using RT and FB techniques (128 and 192 signal averages, respectively; expected scan duration ~13 min each). Spectra were analyzed using jMRUI. SNR and peak areas for PME, Pi, PDE, α‐ATP, and NADPH, normalized using γ‐ATP, were compared. PME/PDE and NADPH/(PME + PDE) ratios were also compared. Statistical Test Paired t ‐tests and Bland–Altman analysis were used. A p value < 0.05 was considered significant.
Results
Scan duration was significantly longer for RT (15 min 44 s) than FB (12 min 56 s). No significant differences were observed for SNR ( p = 0.570), NADPH/(PME + PDE) ( p = 0.931), PME/γ‐ATP ( p = 0.556), Pi/γ‐ATP ( p = 0.931), α‐ATP/γ‐ATP ( p = 0.332), or NADPH/γ‐ATP ( p = 0.394). Significant differences were noted for PDE/γ‐ATP (RT 1.68 vs. FB 1.35, p = 0.003) and PME/PDE (RT 0.434 vs. FB 0.489, p = 0.046). Bland–Altman analysis showed near‐zero fixed biases and no proportional bias, with limits of agreement from −0.53 to 0.62 (PME), −0.30 to 0.30 (Pi), −0.41 to 1.07 (PDE), −0.45 to 0.43 (α‐ATP), and −0.80 to 0.90 (NADPH). Data Conclusion 31 P‐MRS of the liver showed equivalent stability and robustness for RT and FB. FB yielded comparable data within a shorter, predictable scan duration. Evidence Level 2. Technical Efficacy 1.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Hamamatsu University School of Medicine, Hamamatsu, Japan.
- Philips Japan, Tokyo, Japan.
ORCID-Profile: Kumi Ozaki, Jihun Kwon, Marc Van Cauteren, Yasutomo Katsumata
Zitieren
APA
Ozaki, K., Kwon, J., Van Cauteren, M., Katsumata, Y., Tanahashi, Y., & Goshima, S. (2026). Comparison Between Respiratory‐Triggered and Free‐Breathing Signal Acquisition Methods for Single‐Voxel Phosphorus Magnetic Resonance Spectroscopy of the Liver on a Clinical System. Journal of Magnetic Resonance Imaging, 64(3), 702-711. https://doi.org/10.1002/jmri.70352
Vancouver
Ozaki K, Kwon J, Van Cauteren M, Katsumata Y, Tanahashi Y, Goshima S. Comparison Between Respiratory‐Triggered and Free‐Breathing Signal Acquisition Methods for Single‐Voxel Phosphorus Magnetic Resonance Spectroscopy of the Liver on a Clinical System. Journal of Magnetic Resonance Imaging. 2026;64(3):702-711. doi:10.1002/jmri.70352
BibTeX
@article{Ozaki2026,
author = {Ozaki, Kumi and Kwon, Jihun and Van Cauteren, Marc and Katsumata, Yasutomo and Tanahashi, Yukichi and Goshima, Satoshi},
title = {Comparison Between Respiratory‐Triggered and Free‐Breathing Signal Acquisition Methods for Single‐Voxel Phosphorus Magnetic Resonance Spectroscopy of the Liver on a Clinical System},
journal = {Journal of Magnetic Resonance Imaging},
year = {2026},
volume = {64},
number = {3},
pages = {702--711},
doi = {10.1002/jmri.70352}
}
EUVIMED