Multidisciplinary Protocol for 1.5T MRI in Adult Patients With Active Implantable Medical Devices: Safety and Efficacy in a Five‐Year Single‐Center Experience
Journal of Magnetic Resonance Imaging · 2025
DOI: 10.1002/jmri.70188 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Journal of Magnetic Resonance Imaging, 2025. Verfasst von Silvano Filice und 4 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Unit, Azienda Ospedaliero-Universitaria of Parma, Parma, Italy. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Prostheses and Implants. Die Arbeit nennt 45 Literaturangaben.
Abstract (Originalsprache)
Background
The growing population of patients with active implantable medical devices ( AIMDs ) presents challenges for MRI access. Safety concerns regarding device malfunction and patient risk, demand standardized management strategies.
Purpose
To evaluate the safety and feasibility of a structured multidisciplinary safety protocol for 1.5T MRI in AIMD carriers. Study Type Retrospective. Population 514 consecutive patients with AIMDs (cardiac implantable electronic devices (CIEDs) ( n = 347), auditory implants ( n = 85), neurostimulators ( n = 61), and drug delivery systems ( n = 21)) scheduled for clinical MRI (head (34.3%), spine (19.0%), prostate (10.5%), abdomen/pelvis (9.5%), knee (8.6%), heart (3.8%), and other sites (14.3%)) between 2019 and 2025. Field Strength/Sequence MRI was performed on 1.5T scanners using predefined protocols, with automatic compliance to device‐specific SAR limits through the Philips ScanWise Implant software. Assessment The safety protocol included device verification, MRI‐mode programming, compliance with manufacturer safety conditions, physiologic monitoring, and post‐MRI device interrogation. MRI completion rates, exclusion frequencies, and adverse events were recorded, and exclusion proportions were compared across AIMD categories. Statistical Tests Descriptive statistics.
Results
Of 514 patients, 50 (9.7%) with MR ‐unsafe, MR ‐nonconditional, or artifact‐prone devices were not scanned: MR ‐unsafe ( n = 23), MR ‐nonconditional ( n = 22), or anticipated severe artifacts ( n = 3). Four hundred and sixty‐four (90.3%) patients successfully completed diagnostic MRI . Two brain examinations were terminated early due to implant‐site pain. No device resets, malfunctions, or significant changes in sensing/pacing thresholds occurred. Internal magnet displacement was observed in three (3.5%) auditory implants, with two requiring surgical repositioning. Exclusion rates varied by device type, ranging from 11% to 14% for CIEDs to 29% for specific neurostimulators. Data Conclusion A structured, multidisciplinary protocol enables safe MRI in the majority of patients with MR‐conditional AIMDs. Standardized pre‐MRI screening and management support safe implementation, and improve MRI accessibility for this growing and complex patient population. Level of Evidence 4. Technical Efficacy Stage 5.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics Unit, Azienda Ospedaliero-Universitaria of Parma, Parma, Italy.
- Neuroradiology Unit, Azienda Ospedaliero-Universitaria of Parma, Parma, Italy.
- Arrhythmology and Electrophysiology Unit, Azienda Ospedaliero-Universitaria of Parma, Parma, Italy.
- Otorhinolaryngology and Otoneurosurgery Unit, Azienda Ospedaliero-Universitaria of Parma, Parma, Italy.
- Anaesthesia, Intensive Care and Pain Therapy Unit, Azienda Ospedaliero-Universitaria of Parma, Parma, Italy.
ORCID-Profile: Silvano Filice, Maurizio Falcioni
Zitieren
APA
Filice, S., Pavarani, A., Placci, A., Falcioni, M., & Manferdini, M. E. (2025). Multidisciplinary Protocol for 1.5T MRI in Adult Patients With Active Implantable Medical Devices: Safety and Efficacy in a Five‐Year Single‐Center Experience. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70188
Vancouver
Filice S, Pavarani A, Placci A, Falcioni M, Manferdini ME. Multidisciplinary Protocol for 1.5T MRI in Adult Patients With Active Implantable Medical Devices: Safety and Efficacy in a Five‐Year Single‐Center Experience. Journal of Magnetic Resonance Imaging. 2025. doi:10.1002/jmri.70188
BibTeX
@article{Filice2025,
author = {Filice, Silvano and Pavarani, Antonio and Placci, Angelo and Falcioni, Maurizio and Manferdini, Maria Elena},
title = {Multidisciplinary Protocol for 1.5T MRI in Adult Patients With Active Implantable Medical Devices: Safety and Efficacy in a Five‐Year Single‐Center Experience},
journal = {Journal of Magnetic Resonance Imaging},
year = {2025},
doi = {10.1002/jmri.70188}
}
EUVIMED