Quantitative Confounder Analysis of Electrocardiogram Signals in Cardiac Magnetic Resonance at 1.5, 3 and 7 T—Assessing Standardized Electrode Positions and Sequence Types—Towards Quality Assurance
Journal of Magnetic Resonance Imaging · 2025
DOI: 10.1002/jmri.70130 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2025. Verfasst von Richard Hickstein und 10 weiteren Autor:innen. Beteiligte Einrichtung: Working Group Cardiovascular Magnetic Resonance, Experimental and Clinical Research Center, Charité Medical Faculty, Max-Delbrück Center for Molecula…. Hauptthemen laut MEDLINE: Electrocardiography, Heart, Magnetic Resonance Imaging, Cine, Cardiac-Gated Imaging Techniques. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Background
The electrocardiogram (ECG) used for gating in cardiac MRI may be compromised by multiple confounders inside the scanner bore.
Purpose
To quantify the influence of magnetic field strengths (1.5 T/3 T/7 T), standardized electrode positions, and imaging sequences on ECG signals used for gating. Study Type Prospective. Population Sixteen healthy volunteers (eight male; mean age 26.25 ± 7.67 years). Field Strength/Sequence Balanced steady‐state free precession cine (1.5 T/3 T), fast low‐angle shot cine (7 T), and 4D flow (1.5 T/3 T/7 T) sequences. Assessment ECG‐signals were recorded during breath‐hold and non‐breath‐hold short axis cine (sax‐bh and sax‐nbh, respectively) and 4D flow scans at 1.5 T/3 T/7 T. All scans were repeated with 4 standardized electrode positionings (pos1–4) at each field strength. Pos1/2 were vendor‐recommended positionings for 1.5 T/3 T/7 T scans, respectively, whereas pos3/4 were alternative positionings recommended in previous studies. Similarity between confounded ECG‐signals and unconfounded baseline ECG‐signals was assessed by QRS‐feature correlation. Cine image quality (IQ) was assessed by 3 readers (with 6, 10, and 22 years experience) on a four‐point Likert scale. Statistical Tests Linear mixed models with type III tests of fixed effects (overall) and t tests with adjusted degrees of freedom (pairwise subgroup‐comparisons) at significance level p < 0.05.
Results
Increasing field strength resulted in significantly decreasing similarity to baseline measurements, with r values (provided with 95% confidence interval) of 1.5 T: 97% (92.6–101.3); 3 T: 91.4% (87.1–95.8); 7 T: 50.4% (46–54.9) and lower IQ: 1.5 T: 2.33 (2.12–2.55); 3 T: 1.96 (1.75–2.17); 7 T: 0.91 (0.7–1.12). Vendor‐specified electrode positions pos1: 91.8% (87.2–96.5), pos2: 88.3% (83.7–92.9) showed significantly higher correlation with baseline measurements than alternative positions pos3: 67.5% (62.9–72.1) and pos4: 70.8% (66.2–75.4). The evaluated standardized sequences showed similar amounts of electrocardiogram distortion, with r values of: sax‐bh: 77.3% (73–81.7); 4D: 79.3% (75–83.7), p = 0.54; sax‐nbh: 82.1% (77.8–86.5), p = 0.31, but the difference between sax‐bh and sax‐nbh: 4.8% (2.88–6.72) was significant. Data Conclusion Increasing field strength leads to significant ECG signal distortions. Vendor‐specified positions 1/2 resulted in less distorted ECG signals than alternative positions 3/4 recommended in previous publications. Level of Evidence 2. Technical Efficacy Stage 5.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Working Group Cardiovascular Magnetic Resonance, Experimental and Clinical Research Center, Charité Medical Faculty, Max-Delbrück Center for Molecular Medicine, Helios Klinikum Berlin Buch, Department of Cardiology and…
- DZHK (German Centre for Cardiovascular Research), Berlin, Germany.
- Siemens Healthineers AG, Forchheim, Germany.
- Advanta, Siemens SRL, Brasov, Romania.
- Physikalisch-Technische Bundesanstalt (PTB), Berlin, Germany.
ORCID-Profile: Richard Hickstein, Darian Viezzer, Bogdan Andrei Gheorghita, Thomas Hadler, Sebastian Schmitter, Jeanette Schulz-Menger
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Hickstein, R., Wiesemann, S., Viezzer, D., Kleindienst, D., Chitiboi, T., Gheorghita, B. A., Wetzl, J., Hadler, T., Dietrich, S., Schmitter, S., & Schulz-Menger, J. (2025). Quantitative Confounder Analysis of Electrocardiogram Signals in Cardiac Magnetic Resonance at 1.5, 3 and 7 T—Assessing Standardized Electrode Positions and Sequence Types—Towards Quality Assurance. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70130
Vancouver
Hickstein R, Wiesemann S, Viezzer D, Kleindienst D, Chitiboi T, Gheorghita BA, et al. Quantitative Confounder Analysis of Electrocardiogram Signals in Cardiac Magnetic Resonance at 1.5, 3 and 7 T—Assessing Standardized Electrode Positions and Sequence Types—Towards Quality Assurance. Journal of Magnetic Resonance Imaging. 2025. doi:10.1002/jmri.70130
BibTeX
@article{Hickstein2025,
author = {Hickstein, Richard and Wiesemann, Stephanie and Viezzer, Darian and Kleindienst, Denise and Chitiboi, Teodora and Gheorghita, Bogdan Andrei and Wetzl, Jens and Hadler, Thomas and Dietrich, Sebastian and Schmitter, Sebastian and Schulz-Menger, Jeanette},
title = {Quantitative Confounder Analysis of Electrocardiogram Signals in Cardiac Magnetic Resonance at 1.5, 3 and 7 T—Assessing Standardized Electrode Positions and Sequence Types—Towards Quality Assurance},
journal = {Journal of Magnetic Resonance Imaging},
year = {2025},
doi = {10.1002/jmri.70130}
}
EUVIMED