Evaluation of Software‐Optimized Protocols for Acoustic Noise Reduction During Brain MRI at 7 Tesla
Journal of Magnetic Resonance Imaging · 2025
DOI: 10.1002/jmri.29749 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2025. Verfasst von Anton Glans und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nursing, Umeå University, Umeå, Sweden. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Noise, Software. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Region Skåne, Vetenskapsrådet, Swedish governmental funding of clinical research (ALF).
Abstract (Originalsprache)
Background
MR‐generated acoustic noise may be particularly concerning at 7‐Tesla (T) systems. Noise levels can be reduced by altering gradient output using software optimization. However, such alterations might influence image quality or prolong scan times, and these optimizations have not been well characterized.
Purpose
To evaluate image quality, sound pressure levels (SPLs), and perceived noise levels when using the acoustic noise reduction technique SofTone for T 2 ‐weighted fast spin echo (T 2 W FSE) and three‐dimensional T 1 ‐weighted turbo field echo (3D T 1 W TFE), and to compare with conventional imaging during 7‐T brain MRI. Study Type Prospective. Subjects Twenty‐eight volunteers underwent brain MRI, with n = 26 for image quality evaluations. Field Strength/Sequence Conventional and SofTone versions of T 2 W FSE and 3D T 1 W TFE at 7 T. Assessment Peak SPLs (A‐weighted decibels, dBA), participant‐perceived noise levels (Borg CR10‐scale), qualitative image assessments by three neuroradiologists (four‐graded ordinal scales), interrater reliability, and percentage agreement. Statistical Test Paired t ‐test, Wilcoxon's Signed‐Rank Test, and Krippendorff's alpha; p < 0.05 were considered statistically significant.
Results
SofTone significantly reduced peak SPLs: from 116.3 to 97.0 dBA on T 2 W FSE, and from 123.7 to 101.5 dBA on 3D T 1 W TFE. SofTone was perceived as significantly quieter than conventional scanning. T 2 W FSE showed no significant differences in image quality assessments ( p = 0.21–1.00), except one radiologist noting significantly less artifact interference with SofTone. General image quality remained acceptable for 3D T 1 W TFE, though one radiologist scored it significantly lower with SofTone (mean scores: 3.08 vs. 3.65), and two radiologists observed significantly worse white and gray matter differentiation with SofTone (mean scores: 3.19 vs. 3.54; 2.27 vs. 2.81). Data Conclusion SofTone can significantly reduce sound intensity and perceived noise levels while maintaining acceptable image quality with T 2 W FSE and 3D T 1 W TFE in brain MRI. It appears to be an effective tool for providing a safer, quieter 7‐T scan environment. Evidence Level: 4 Technical Efficacy: Stage 5
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Nursing, Umeå University, Umeå, Sweden.
- Department of Diagnostics and Intervention, Radiation Physics, Umeå University, Umeå, Sweden.
- Department of Clinical Sciences Lund, Diagnostic Radiology, Faculty of Medicine, Lund University, Lund, Sweden.
- Department of Medical Imaging and Physiology, Skåne University Hospital, Lund, Sweden.
- Lund BioImaging Centre, Faculty of Medicine, Lund University, Lund, Sweden.
- Department of Clinical Sciences Lund, Logopedics, Phoniatrics and Audiology, Faculty of Medicine, Lund University, Lund, Sweden.
ORCID-Profile: Anton Glans, Boel Hansson
Förderung
- Region Skåne
- Vetenskapsrådet
- Swedish governmental funding of clinical research (ALF)
Zitieren
APA
Glans, A., Wennberg, L., Wilén, J., Lindgren, L., Sundgren, P. C., Mårtensson, J., Markenroth Bloch, K., & Hansson, B. (2025). Evaluation of Software‐Optimized Protocols for Acoustic Noise Reduction During Brain MRI at 7 Tesla. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29749
Vancouver
Glans A, Wennberg L, Wilén J, Lindgren L, Sundgren PC, Mårtensson J, et al. Evaluation of Software‐Optimized Protocols for Acoustic Noise Reduction During Brain MRI at 7 Tesla. Journal of Magnetic Resonance Imaging. 2025. doi:10.1002/jmri.29749
BibTeX
@article{Glans2025,
author = {Glans, Anton and Wennberg, Linda and Wilén, Jonna and Lindgren, Lenita and Sundgren, Pia C and Mårtensson, Johan and Markenroth Bloch, Karin and Hansson, Boel},
title = {Evaluation of Software‐Optimized Protocols for Acoustic Noise Reduction During Brain MRI at 7 Tesla},
journal = {Journal of Magnetic Resonance Imaging},
year = {2025},
doi = {10.1002/jmri.29749}
}
EUVIMED