Robustness of Quantitative Diffusion Metrics from Four Models: A Prospective Study on the Influence of Scan‐Rescans, Voxel Size, Coils, and Observers

Jingyu Zhong iD, Xianwei Liu, Yangfan Hu, Yue Xing, Defang Ding, Xiang Ge, Yang Song iD, Silian Wang, Liwei Chen, Ying Zhu, Wenjie Lu, Huan Zhang iD und 1 weitere

Journal of Magnetic Resonance Imaging · 2024

DOI: 10.1002/jmri.29192 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2024. Verfasst von Jingyu Zhong und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China. Hauptthemen laut MEDLINE: Brain, Diffusion Magnetic Resonance Imaging, Observer Variation. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Natural Science Foundation of China, Guangci Innovative Technology Launch Plan of Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Research Fund of Tongren Hospital, Shanghai Jiao Tong University School of Medicine.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Quantitative diffusion metrics provide additional microstructural information of diseases. The robustness of quantitative diffusion metrics should be established before clinical application.

Purpose

To evaluate the variability and reproducibility of quantitative diffusion MRI metrics. Study Type Prospective. Population 14 volunteers (7 men; median age, range, 28, 26–59 years). Field strength/Sequence 3. 0‐T /Diffusion spectrum imaging. Assessment Brain MRI studies were performed four times per subject: involving different combinations of coil types and voxel sizes. Regions of interest of 13 brain anatomical sites were drawn by one observer twice and another observer once to allow interobserver and intraobserver reproducibility assessment. Twenty‐five quantitative metrics were calculated using four diffusion models. Statistical Tests The variability was evaluated with coefficients of variation (CV), and quartile coefficient of dispersion (QCD). The reproducibility was assessed with intraclass correlation coefficient (ICC), and concordance correlation coefficient (CCC). Wilcoxon signed rank test was used to compare the influence of factors on robustness of quantitative diffusion metrics. A two‐tailed P < 0.05 was considered statistically significant.

Results

The variability of quantitative diffusion metrics showed CV of 2.4%–68.2%, and QCD of 0.6%–48.2%, respectively. The reproducibility of scans using 20‐channel coils with voxels of 2 × 2 × 2 mm 3 and 3 × 3 × 3 mm 3 , respectively (ICC 0.03–0.84, CCC 0.03–0.84) was significantly worse than that of repeated scans using a 20‐channel coil with a voxel size of 2 × 2 × 2 mm 3 (ICC of 0.74–0.97, CCC 0.74–0.97) and that of scans using 20‐ and 64‐channel coils, respectively, with a voxel size of 2 × 2 × 2 mm 3 (ICC 0.59–0.95, CCC 0.59–0.95). The intraobserver reproducibility (ICC 0.49–0.94, CCC 0.49–0.94) was significantly better than the interobserver reproducibility (ICC 0.28–0.91, CCC 0.28–0.91). Data Conclusion Our study indicated that the voxel size has a greater influence on the reproducibility of quantitative diffusion metrics than scan‐rescans and coils. The reproducibility within one observer was higher than that between two observers. Evidence Level 2 Technical Efficacy Stage 1

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrainDiffusion Magnetic Resonance ImagingObserver VariationHumansProspective StudiesReproducibility of ResultsAdultFemaleMaleMiddle AgedImage Processing, Computer-AssistedImage Interpretation, Computer-AssistedAlgorithms

Stichwörter der Autor:innen

braindiffusion magnetic resonance imagingreproducibility of results

Autor:innen und Institutionen

  • Department of Imaging, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • MR Scientific Marketing, Siemens Healthineers Ltd, Shanghai, China.
  • Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.

ORCID-Profile: Jingyu Zhong, Yang Song, Huan Zhang

Förderung

  • National Natural Science Foundation of China
  • Guangci Innovative Technology Launch Plan of Ruijin Hospital, Shanghai Jiao Tong University School of Medicine
  • Research Fund of Tongren Hospital, Shanghai Jiao Tong University School of Medicine
  • Yangfan Project of Science and Technology Commission of Shanghai Municipality

Zitieren

APA

Zhong, J., Liu, X., Hu, Y., Xing, Y., Ding, D., Ge, X., Song, Y., Wang, S., Chen, L., Zhu, Y., Lu, W., Zhang, H., & Yao, W. (2024). Robustness of Quantitative Diffusion Metrics from Four Models: A Prospective Study on the Influence of Scan‐Rescans, Voxel Size, Coils, and Observers. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29192

Vancouver

Zhong J, Liu X, Hu Y, Xing Y, Ding D, Ge X, et al. Robustness of Quantitative Diffusion Metrics from Four Models: A Prospective Study on the Influence of Scan‐Rescans, Voxel Size, Coils, and Observers. Journal of Magnetic Resonance Imaging. 2024. doi:10.1002/jmri.29192

BibTeX
@article{Zhong2024,
  author = {Zhong, Jingyu and Liu, Xianwei and Hu, Yangfan and Xing, Yue and Ding, Defang and Ge, Xiang and Song, Yang and Wang, Silian and Chen, Liwei and Zhu, Ying and Lu, Wenjie and Zhang, Huan and Yao, Weiwu},
  title = {Robustness of Quantitative Diffusion Metrics from Four Models: A Prospective Study on the Influence of Scan‐Rescans, Voxel Size, Coils, and Observers},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2024},
  doi = {10.1002/jmri.29192}
}
Kostenlos starten →