Effect of design and surgical parameters variations in mobile‐bearing versus fixed‐bearing unicompartmental knee arthroplasty: A finite element analysis
Journal of Experimental Orthopaedics · 2024-10 · 11 (4)
DOI: 10.1002/jeo2.70053 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Experimental Orthopaedics, online erschienen am 26.10.2024. Verfasst von Thomas Luyckx und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery AZ Delta Roeselare Belgium. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 74 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Unicompartmental knee arthroplasty (UKAs) are available in the market as fixed‐ and mobile‐bearing (FB and MB) and can be characterised by a different set of design parameters in terms of geometries, materials and surgical approaches, with overall good clinical outcomes. However, clear biomechanical evidence concerning the consequences of variations of these features on knee biomechanics is still lacking; therefore, the present study aims to perform a sensitivity analysis to see which outcomes are affected by these variations.
Methods
For both MB‐UKA and FB‐UKA, five design and surgical parameters were defined (bearing insert thickness, tibial component material, implant components friction coefficient, antero‐posterior slope angle and level of tibial bone resection). Two control models were defined based on standard configurations for both implants. Finite element analysis was chosen to perform this study, and different parameter combinations (216 models in total) were implemented and tested at both 0° and 90° of flexion, using a previously validated finite element knee model. The results were then evaluated in terms of bone and polyethylene Von Mises stress and tibio‐femoral contact area.
Results
Bearing thickness, tibial bone cut and slope angle were found to be the most sensitive parameters for both types of UKAs. Specifically, changes in these parameters in the FB‐UKA appeared to induce more significant variations in the polyethylene insert (both in terms of polyethylene stress and contact area), while in the MB‐UKA, these changes influenced bone stress distribution more.
Conclusions
Surgical parameters returned to have a more significant influence than material and friction variations; furthermore, the outcomes most affected by parameter variations were the insert‐related ones for FB‐UKA while for the MB‐UKA were the ones regarding tibial bone stresses. Level of Evidence Not Applicable.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedic Surgery AZ Delta Roeselare Belgium.
- BEAMS Department (Bio Electro and Mechanical Systems), École Polytechnique de Bruxelles Université Libre de Bruxelles Bruxelles Belgium.
ORCID-Profile: Thomas Luyckx, Edoardo Bori, Rachele Saldari, Bernardo Innocenti
Zitieren
APA
Luyckx, T., Bori, E., Saldari, R., Fiore, S., Altamore, V., & Innocenti, B. (2024). Effect of design and surgical parameters variations in mobile‐bearing versus fixed‐bearing unicompartmental knee arthroplasty: A finite element analysis. Journal of Experimental Orthopaedics, 11(4). https://doi.org/10.1002/jeo2.70053
Vancouver
Luyckx T, Bori E, Saldari R, Fiore S, Altamore V, Innocenti B. Effect of design and surgical parameters variations in mobile‐bearing versus fixed‐bearing unicompartmental knee arthroplasty: A finite element analysis. Journal of Experimental Orthopaedics. 2024;11(4). doi:10.1002/jeo2.70053
BibTeX
@article{Luyckx2024,
author = {Luyckx, Thomas and Bori, Edoardo and Saldari, Rachele and Fiore, Sara and Altamore, Virginia and Innocenti, Bernardo},
title = {Effect of design and surgical parameters variations in mobile‐bearing versus fixed‐bearing unicompartmental knee arthroplasty: A finite element analysis},
journal = {Journal of Experimental Orthopaedics},
year = {2024},
volume = {11},
number = {4},
doi = {10.1002/jeo2.70053}
}
EUVIMED