Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System

Jia Liu iD, Lin Wang iD, Yue Tang iD, Peiran Li iD, Lili Zhao, Wenwen Yuan iD, Bo Li iD, Boan Li iD

Journal of Clinical Laboratory Analysis · 2026-11 · 40 (21)

DOI: 10.1002/jcla.70347 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, online erschienen am 14.10.2026. Verfasst von Jia Liu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory, The Fifth Medical Center, Chinese People's Liberation Army General Hospital, Beijing, China. Hauptthemen laut MEDLINE: Ammonia, Colorimetry, Blood Chemical Analysis. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Plasma ammonia quantitation is one of the most valuable laboratory tests for assessing hepatic function. Accurate measurement is clinically imperative for the diagnosis and management of severe hepatobiliary diseases. This study aimed to comprehensively evaluate the analytical performance of the VITROS dry chemistry system with dry chemical colorimetry for plasma ammonia detection.

Methods

A full performance validation was conducted on the VITROS 350 Chemistry System in accordance with the guidelines of the Clinical and Laboratory Standards Institute (CLSI) and the China National Accreditation Service for Conformity Assessment (CNAS). Comparability was performed using the VITROS 350 Chemistry System and VITROS 5600 Integrated System across our center and four external laboratories.

Results

The VITROS 350 platform demonstrated robust analytical performance for plasma ammonia testing. Precision, accuracy, upper limit of the reportable interval, linearity, and carryover rates were all compliant with both clinical requirements and manufacturer specifications. Comparative analysis revealed strong inter‐platform concordance, with results from five concentration levels across five laboratories remaining within predefined allowable limits. Regarding interference, only high concentrations of hemoglobin exerted a clinically significant impact on test results, a finding consistent with established clinical experience and diagnostic expectations.

Conclusion

This study validates the reliability of the VITROS dry chemistry system for routine plasma ammonia testing. It provides clinical laboratories with a robust framework for method selection and establishes a framework for future external quality assessment (EQA) based on dry chemistry colorimetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AmmoniaColorimetryBlood Chemical AnalysisHumansReproducibility of ResultsLimit of Detection

Stichwörter der Autor:innen

ammoniaanalytical performancedry chemical colorimetryinterferencetest system

Substanzen

Ammonia

Autor:innen und Institutionen

  • Department of Clinical Laboratory, The Fifth Medical Center, Chinese People's Liberation Army General Hospital, Beijing, China.
  • Chinese People's Liberation Army Center for Disease Control and Prevention, Beijing, China.

ORCID-Profile: Jia Liu, Lin Wang, Yue Tang, Peiran Li, Wenwen Yuan, Bo Li, Boan Li

Zitieren

APA

Liu, J., Wang, L., Tang, Y., Li, P., Zhao, L., Yuan, W., Li, B., & Li, B. (2026). Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System. Journal of Clinical Laboratory Analysis, 40(21). https://doi.org/10.1002/jcla.70347

Vancouver

Liu J, Wang L, Tang Y, Li P, Zhao L, Yuan W, et al. Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System. Journal of Clinical Laboratory Analysis. 2026;40(21). doi:10.1002/jcla.70347

BibTeX
@article{Liu2026,
  author = {Liu, Jia and Wang, Lin and Tang, Yue and Li, Peiran and Zhao, Lili and Yuan, Wenwen and Li, Bo and Li, Boan},
  title = {Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2026},
  volume = {40},
  number = {21},
  doi = {10.1002/jcla.70347}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche