The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900
Journal of Clinical Laboratory Analysis · 2024-04 · 38 (8)
DOI: 10.1002/jcla.25007 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, online erschienen am 09.04.2024. Verfasst von Filipa P Freitas und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal. Hauptthemen laut MEDLINE: Leukocytes. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
The Beckman Coulter DxH 900 is a haematological analyser capable of counting and sizing blood cells, and obtaining a complete blood cell count (CBC). This analyses different parameters of red blood cells (RBC), platelets and white blood cells/leukocytes. Some automated CBC counters present limitations due to specimen characteristics, abnormal cells or both factors. In the presence of abnormalities, the DxH 900 has a flagging system, warning the laboratory technician that something needs to be verified. In the present work, we evaluated samples from oncologic patients, presenting a population erroneously perceived as being lymphocytes. The most common explanations for this situation are RBC resistant to lysis or serum hyperbilirubinaemia.
Methods
In an attempt to solve and understand what the cause of this problem might be, we diluted our samples (1:3) and analysed the serum total bilirubin. To identify cells' abnormalities, the samples were also analysed by manual DLC counts. During the study, we also checked the different flags presented by the equipment.
Results
The results evidenced that the major interference was due to RBC lysis resistance, corresponding to 94.7% of the cases, while hyperbilirubinaemia was only present in 73.4%. Besides, we determined that some samples with normal bilirubin levels also presented interference, suggesting that hyperbilirubinaemia was not the main cause of the error. The most recurrent flag observed was “High event rate”.
Conclusion
The dilution solved all of the observed interferences. The results between diluted and manual counts showed a strong correlation, leading us to introduce dilution in our laboratory routine.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Clinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal.
- University of Maia, Maia, Portugal.
- NECE - Research Center for Business Sciences, Universidade da Beira Interior, Covilhã, Portugal.
- Institute of Medical Microbiology, Faculty of Medicine, University of Coimbra, Coimbra, Portugal.
ORCID-Profile: Joana Oliveira
Zitieren
APA
Freitas, F. P., Reis, J., Oliveira, J., Veiga, P. M., Raquel Paiva, A., & Soares, R. (2024). The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900. Journal of Clinical Laboratory Analysis, 38(8). https://doi.org/10.1002/jcla.25007
Vancouver
Freitas FP, Reis J, Oliveira J, Veiga PM, Raquel Paiva A, Soares R. The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900. Journal of Clinical Laboratory Analysis. 2024;38(8). doi:10.1002/jcla.25007
BibTeX
@article{Freitas2024,
author = {Freitas, Filipa P and Reis, Jorge and Oliveira, Joana and Veiga, Pedro Mota and Raquel Paiva, Ana and Soares, Rui},
title = {The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900},
journal = {Journal of Clinical Laboratory Analysis},
year = {2024},
volume = {38},
number = {8},
doi = {10.1002/jcla.25007}
}
EUVIMED