Fachartikel in der Zeitschrift Journal of Biomedical Materials Research Part B: Applied Biomaterials, online erschienen am 19.01.2026. Verfasst von Maximilien E Launey und 8 weiteren Autor:innen. Beteiligte Einrichtung: G.RAU Inc., Scotts Valley, California, USA. Hauptthemen laut MEDLINE: Alloys, Materials Testing, Heart Valve Prosthesis, Transcatheter Mitral Valve Replacement. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Transcatheter cardiovascular devices require high‐purity Nitinol materials with exceptional fatigue resistance to meet stringent Class III regulatory durability requirements. Ultra‐clean VAR/EBR (Vacuum Arc Remelt/Electron Beam Remelt) Nitinol represents a metallurgical advancement that achieves unprecedented control over inclusion size and distribution. This study characterizes the fatigue behavior of VAR/EBR Nitinol with nominal inclusion sizes below 10 μm under conditions representative of transcatheter mitral valve replacement (TMVR) applications, using the HighLife Mitral Valve Replacement system as a clinical case study. Diamond‐shaped fatigue specimens were manufactured from ultra‐clean VAR/EBR Nitinol tubing used in the HighLife Mitral Valve Replacement system and tested under physiologically relevant conditions to 10 7 , 10 8 , and 4 × 10 8 cycles. Testing included multiple combinations of mean strains (1.5%–9%) and strain amplitudes (0.50%–2.50%) to simulate the multi‐strain operating environments encountered in TMVR devices. VAR/EBR Nitinol demonstrated a conservative 130% improvement in 4 × 10 8 ‐cycle Fatigue Strain Limit (FSL) compared to conventional VAR Nitinol at mean strains between 3% and 5%. The FSL behavior revealed two distinct strain regimes correlating with stress‐induced martensitic transformation. Fractographic analysis confirmed elimination of inclusion‐initiated fatigue failure, with crack initiation occurring independently of microstructural defects. The ultra‐clean microstructure of VAR/EBR Nitinol enables a fundamental shift from flaw‐dominated to stress‐dominated fatigue behavior, providing unprecedented safety margins for complex transcatheter devices operating across diverse mechanical conditions. This material advancement has broad implications for Class III cardiovascular device design, enabling devices like the HighLife system and other TMVR platforms to meet stringent regulatory durability requirements while maintaining safety across complex multi‐strain operating environments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlloysMaterials TestingHeart Valve ProsthesisTranscatheter Mitral Valve ReplacementHumansStress, MechanicalProsthesis Design

Stichwörter der Autor:innen

HighLifeVAR/EBRclass III medical devicefatiguenitinoltranscatheter mitral valve

Substanzen

Alloysnitinol

Autor:innen und Institutionen

  • G.RAU Inc., Scotts Valley, California, USA.
  • HighLife Medical Inc./HighLife SAS, Paris, France.
  • Engage Medical Device Services Inc., Newport Beach, California, USA.
  • Euroflex GmbH, Pforzheim, Germany.
  • Department of Materials Science and Engineering, Stanford University, Stanford, California, USA.

ORCID-Profile: Maximilien E Launey

Zitieren

APA

Launey, M. E., Bobo, T., Miri, B., Park, S., Kayo, M., Saffari, P., Pelton, S. M., Ulmer, J., & Pelton, A. R. (2026). Enhanced Fatigue Performance of VAR / EBR Nitinol for Transcatheter Heart Valve Applications: Material Characterization and Design Implications. Journal of Biomedical Materials Research Part B: Applied Biomaterials, 114(1). https://doi.org/10.1002/jbmb.70034

Vancouver

Launey ME, Bobo T, Miri B, Park S, Kayo M, Saffari P, et al. Enhanced Fatigue Performance of VAR / EBR Nitinol for Transcatheter Heart Valve Applications: Material Characterization and Design Implications. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 2026;114(1). doi:10.1002/jbmb.70034

BibTeX
@article{Launey2026,
  author = {Launey, Maximilien E and Bobo, Trey and Miri, Behnood and Park, Steven and Kayo, Margaret and Saffari, Payman and Pelton, Sean M and Ulmer, Jochen and Pelton, Alan R},
  title = {Enhanced Fatigue Performance of VAR / EBR Nitinol for Transcatheter Heart Valve Applications: Material Characterization and Design Implications},
  journal = {Journal of Biomedical Materials Research Part B: Applied Biomaterials},
  year = {2026},
  volume = {114},
  number = {1},
  doi = {10.1002/jbmb.70034}
}