Synergistic Effect of Implant Surface Physicochemical Modifications and Macrogeometry on the Early Stages of Osseointegration: An In Vivo Preclinical Study
Journal of Biomedical Materials Research Part B: Applied Biomaterials · 2025
DOI: 10.1002/jbm.b.35569 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Biomedical Materials Research Part B: Applied Biomaterials, 2025. Verfasst von Ernesto B Benalcázar-Jalkh und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics and Periodontology, Bauru School of Dentistry, University of Sao Paulo, Bauru, Sao Paulo, Brazil. Hauptthemen laut MEDLINE: Osseointegration, Dental Implants, Implants, Experimental. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
This preclinical, in vivo study aimed to histologically and histomorphometrically evaluate the effect of implant design features on bone healing during the early stages of osseointegration. Three different implant macrogeometries and surface treatments were evaluated: (1) trapezoidal threads with decompressing vertical chambers and blasted acid‐etched surface (Maestro/Blasted+AE); (2) large thread pitch implant with deep and wide threads, with TiUnite surface (RS/TiUnite); and (3) progressive buttress threads with SLActive surface (BL/SLActive). Implant surfaces were characterized by scanning electron microscopy, profilometry, and energy‐dispersive X‐ray spectroscopy. Implants were placed in the iliac bone of 12 female sheep (~65 kg and 2 years old). Following healing times of 3‐ and 6‐ weeks, samples were harvested and subjected to qualitative and quantitative histological/histomorphometric evaluations. Percentages of bone‐to‐implant contact (%BIC) along the implant's perimeter and bone area fraction occupancy (%BAFO) within implant threads were measured, and results were analyzed using a linear mixed model analysis. All implants, irrespective of differences in macrogeometry and surface treatment, at both healing times demonstrated successful osseointegration. Evaluations of %BIC yielded no statistically significant differences among groups at 3 and 6 weeks ( p > 0.052). While no significant differences were detected among groups for %BAFO at 3 weeks ( p > 0.249), Maestro/Blasted+AE yielded significantly higher degrees of bone formation within implant threads relative to RS/TiUnite ( p = 0.043) and BL/SLActive group ( p = 0.032) at the 6‐week time point. Qualitative histological analyses depicted different osseointegration features for the different implants. While Maestro/Blasted+AE portrayed evidence of an intramembranous‐like osseointegration pathway in the healing chambers and interfacial remodeling at thread tips, BL/SLActive and RS/TiUnite groups predominantly presented an interfacial bone remodeling healing pathway. Implant design features influenced the osseointegration pathway, where implants with decompressing vertical chambers enhanced bone formation between implant threads.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Prosthodontics and Periodontology, Bauru School of Dentistry, University of Sao Paulo, Bauru, Sao Paulo, Brazil.
- Department of Biochemistry and Molecular Biology, University of Miami Miller School of Medicine, Miami, Florida, USA.
- Dr. John T. Macdonald Foundation Biomedical Nanotechnology Institute (BioNIUM), University of Miami, Miami, Florida, USA.
- University of Miami Miller School of Medicine, Miami, Florida, USA.
- Department of Diagnosis and Surgery, School of Dentistry of Araraquara, Sao Paulo State University, R. Humaitá, Sao Paulo, Brazil.
- DeWitt Daughtry Family Department of Surgery, Division of Plastic Surgery, University of Miami Miller School of Medicine, Miami, Florida, USA.
ORCID-Profile: Ernesto B Benalcázar-Jalkh, Lukasz Witek
Zitieren
APA
Benalcázar-Jalkh, E. B., Nayak, V. V., Slavin, B. V., Balderrama, I. F., Bonfante, E. A., Coelho, P. G., & Witek, L. (2025). Synergistic Effect of Implant Surface Physicochemical Modifications and Macrogeometry on the Early Stages of Osseointegration: An In Vivo Preclinical Study. Journal of Biomedical Materials Research Part B: Applied Biomaterials. https://doi.org/10.1002/jbm.b.35569
Vancouver
Benalcázar-Jalkh EB, Nayak VV, Slavin BV, Balderrama IF, Bonfante EA, Coelho PG, et al. Synergistic Effect of Implant Surface Physicochemical Modifications and Macrogeometry on the Early Stages of Osseointegration: An In Vivo Preclinical Study. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 2025. doi:10.1002/jbm.b.35569
BibTeX
@article{BenalczarJalkh2025,
author = {Benalcázar-Jalkh, Ernesto B and Nayak, Vasudev Vivekanand and Slavin, Blaire V and Balderrama, Isis Fatima and Bonfante, Estevam A and Coelho, Paulo G and Witek, Lukasz},
title = {Synergistic Effect of Implant Surface Physicochemical Modifications and Macrogeometry on the Early Stages of Osseointegration: An In Vivo Preclinical Study},
journal = {Journal of Biomedical Materials Research Part B: Applied Biomaterials},
year = {2025},
doi = {10.1002/jbm.b.35569}
}
EUVIMED