Comparison of Outcomes After Anterior Cervical Discectomy and Fusion Using Bioactive Glass–Ceramic Spacer‐7 ( NOVOMAX ‐ FUSION ) and an Allograft Spacer: A Retrospective, Case‐Matched, Multicenter Study
Journal of Biomedical Materials Research Part B: Applied Biomaterials · 2025
DOI: 10.1002/jbm.b.35521 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Biomedical Materials Research Part B: Applied Biomaterials, 2025. Verfasst von Gi-Wuk Jang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Spine and Spinal Cord Institute, Gangnam Severance Spine Hospital, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Ceramics, Spinal Fusion, Diskectomy, Cervical Vertebrae. Die Arbeit nennt 34 Literaturangaben.
Abstract (Originalsprache)
The bioactive glass–ceramic spacer (BGS)‐7, a biosynthetic intervertebral fusion material introduced in 2014, has not been the subject of comparative clinical studies on anterior cervical discectomy and fusion (ACDF) surgery. This study, for the first time, aims to compare the radiological and clinical outcomes of the renewed BGS‐7, released in 2019, with those of an allograft spacer. The comparison includes a finite element analysis of the biomechanical properties of each implant, adding a novel dimension to the research. We prospectively followed up on 29 patients who underwent ACDF using BGS‐7 as the experimental group. To select a control group for comparison, 253 patients with level 1 ACDF with an allograft spacer between 2012 and 2022 were selected from our hospital. Using propensity score matching, 27 and 54 patients in the BGS‐7 and allograft groups, respectively, were selected. The average subsidence length was 1.02 ± 1.44 mm per level in the BGS‐7 group and 2.27 ± 2.25 mm per level in the allograft group. Subsidence was observed in 14 of 54 patients (25.9%) in the allograft group and one of 27 patients (3.7%) in the BGS‐7 group ( p = 0.016). In the allograft group, 16 of the 54 patients (29.6%) monitored for 6 months achieved satisfactory fusion outcomes with grades 4 and 5. Thirty‐eight of 54 patients (70.4%) followed up for > 1 year in the allograft group achieved adequate fusion outcomes with grades 4 and 5. In the BGS‐7 group, 17 of the 27 patients (63.0%) monitored for 6 months achieved satisfactory fusion results with grades 4 and 5. Twenty‐three of the 27 patients (85.2%) followed up for > 1 year obtained adequate fusion outcomes with grades 4 and 5. There was a significant difference in the fusion rates between the two groups at 6 months ( p = 0.008). BGS‐7 is a reliable instrument for ACDF with no instances of instrumental failure. The BGS‐7 group had positive clinical outcomes after surgery without any untoward events, and an early fusion rate with the creation of a bone bridge was noted during the 6‐month follow‐up period. Our findings not only indicate the safety of BGS‐7 but also its practicality as a substitute for allografts in ACDF, instilling confidence in its application.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Neurosurgery, Spine and Spinal Cord Institute, Gangnam Severance Spine Hospital, Seoul, Republic of Korea.
- Department of Neurosurgery, Yonsei University College of Medicine, Seoul, Republic of Korea.
- Department of Neurosurgery, Severance Hospital, Seoul, Republic of Korea.
ORCID-Profile: Gi-Wuk Jang
Zitieren
APA
Jang, G. W., Kim, K. H., Shin, D. A., Jang, H. J., Lee, C. K., Chin, D. K., Park, J. Y., Koo, S. W., & Moon, B. J. (2025). Comparison of Outcomes After Anterior Cervical Discectomy and Fusion Using Bioactive Glass–Ceramic Spacer‐7 ( NOVOMAX ‐ FUSION ) and an Allograft Spacer: A Retrospective, Case‐Matched, Multicenter Study. Journal of Biomedical Materials Research Part B: Applied Biomaterials. https://doi.org/10.1002/jbm.b.35521
Vancouver
Jang GW, Kim KH, Shin DA, Jang HJ, Lee CK, Chin DK, et al. Comparison of Outcomes After Anterior Cervical Discectomy and Fusion Using Bioactive Glass–Ceramic Spacer‐7 ( NOVOMAX ‐ FUSION ) and an Allograft Spacer: A Retrospective, Case‐Matched, Multicenter Study. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 2025. doi:10.1002/jbm.b.35521
BibTeX
@article{Jang2025,
author = {Jang, Gi-Wuk and Kim, Kyung-Hyun and Shin, Dong-Ah and Jang, Hyun-Joon and Lee, Chang-Kyu and Chin, Dong-Kyu and Park, Jeong-Yoon and Koo, Seong-Wook and Moon, Bong-Ju},
title = {Comparison of Outcomes After Anterior Cervical Discectomy and Fusion Using Bioactive Glass–Ceramic Spacer‐7 ( NOVOMAX ‐ FUSION ) and an Allograft Spacer: A Retrospective, Case‐Matched, Multicenter Study},
journal = {Journal of Biomedical Materials Research Part B: Applied Biomaterials},
year = {2025},
doi = {10.1002/jbm.b.35521}
}
EUVIMED