EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Increased stability due to symmetric cement volume in augmented pedicle screws? A biomechanical study

Maximilian Weber iD, Maximilian Lenz, Philipp Egenolf, Andreas Prescher, Sebastian Walter, Vincent J Heck, Peer Eysel, Max Joseph Scheyerer

Journal of Biomedical Materials Research Part B: Applied Biomaterials · 2024-01 · 112 (1)

DOI: 10.1002/jbm.b.35339 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Biomedical Materials Research Part B: Applied Biomaterials, online erschienen am 13.11.2023. Verfasst von Maximilian Weber und 7 weiteren Autor:innen. Beteiligte Einrichtung: Center for Orthopedic and Trauma Surgery, Faculty of Medicine and University Hospital, University of Cologne, Cologne, Germany. Hauptthemen laut MEDLINE: Pedicle Screws. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pedicle screw instrumentation has become “state of the art” in surgical treatment of many spinal disorders. Loosening of pedicle screws due to poor bone mineral density is a frequent complication in osteoporotic patients. As prevalence of osteoporosis and spinal disorders are increasing with an aging demographic, optimizing the biomechanical properties of pedicle screw constructions and therefore outcome after spinal surgery in osteoporotic patients is a key factor in future surgical therapy. Therefore, this biomechanical study investigated the stability of polymethylmethacrylate (PMMA)‐augmented pedicle screw‐rod constructions under a deviating distribution of PMMA applied to the instrumentation in osteoporotic human cadaveric vertebrae. We showed that PMMA‐augmented pedicle screw‐rod constructions tend to be more stable than those with non‐augmented pedicle screws. Further, there appears to be a larger risk of screw loosening in unilateral augmented pedicle screws than in non‐augmented, therefore a highly asymmetrically distributed PMMA should be avoided.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pedicle ScrewsHumansPolymethyl MethacrylateLumbar VertebraeBiomechanical PhenomenaBone Cements

Stichwörter der Autor:innen

PMMAbiomechanicsosteoporosispedicle screwscrew-related complicationsspine surgery

Substanzen

Polymethyl MethacrylateBone Cements

Autor:innen und Institutionen

  • Center for Orthopedic and Trauma Surgery, Faculty of Medicine and University Hospital, University of Cologne, Cologne, Germany.
  • University Hospital RWTH, Institute of Molecular and Cellular Anatomy (MOCA), Aachen, Germany.
  • Department of Trauma and Spine Surgery, Heinrich Heine University Hospital Düsseldorf, Düsseldorf, Germany.

ORCID-Profile: Maximilian Weber

Zitieren

APA

Weber, M., Lenz, M., Egenolf, P., Prescher, A., Walter, S., Heck, V. J., Eysel, P., & Scheyerer, M. J. (2024). Increased stability due to symmetric cement volume in augmented pedicle screws? A biomechanical study. Journal of Biomedical Materials Research Part B: Applied Biomaterials, 112(1). https://doi.org/10.1002/jbm.b.35339

Vancouver

Weber M, Lenz M, Egenolf P, Prescher A, Walter S, Heck VJ, et al. Increased stability due to symmetric cement volume in augmented pedicle screws? A biomechanical study. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 2024;112(1). doi:10.1002/jbm.b.35339

BibTeX
@article{Weber2024,
  author = {Weber, Maximilian and Lenz, Maximilian and Egenolf, Philipp and Prescher, Andreas and Walter, Sebastian and Heck, Vincent J and Eysel, Peer and Scheyerer, Max Joseph},
  title = {Increased stability due to symmetric cement volume in augmented pedicle screws? A biomechanical study},
  journal = {Journal of Biomedical Materials Research Part B: Applied Biomaterials},
  year = {2024},
  volume = {112},
  number = {1},
  doi = {10.1002/jbm.b.35339}
}
Kostenlos starten →