The anticorrosion ability of titanium nitride (TiN) plating on an orthodontic metal bracket and its biocompatibility
Journal of Biomedical Materials Research · 2002-01 · 63 (6) · 786-792
DOI: 10.1002/jbm.10484 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Biomedical Materials Research, online erschienen am 18.10.2002. Verfasst von Chia-Tze Kao und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Stomatology, Chung Shan Medical and Dental College, 110, Section 1, Chien Kuo N Road, Taichung, Taiwan, PO 402. Hauptthemen laut MEDLINE: Coated Materials, Biocompatible, Orthodontic Brackets, Titanium. Die Arbeit wurde bisher 61-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Typically, an orthodontic metal bracket is made from stainless steel. It has been shown that such metal may corrode in an acid‐ and chloride‐rich environment. The purpose of the current study was to investigate a titanium nitride (TiN) ion‐plated stainless steel orthodontic bracket's anticorrosion properties and compare its biocompatibility with that of non‐TiN‐plated brackets. The stainless‐steel brackets studied here were tested in acidic artificial saliva. The plated metal bracket was produced by the titanium nitride (TiN) ion‐plating method. The TiN‐plating on the bracket surface was demonstrated to be successful by EDX analysis. The quantity of metallic‐ion release under test immersion solutions was analyzed by atomic absorption spectrophotometry. Both TiN‐ and non‐TiN‐plated brackets may release detectable ions into the test solution, including nickel, chromium, manganese, copper, and iron (ferric). The anticorrosion ability of the plated bracket was analyzed by means of inductively coupled plasma atomic emission spectroscopy. The results revealed that the TiN‐plated metal bracket did not increase the anticorrosion ability of the standard bracket. The biocompatibility of the TiN plating versus the standard bracket material resulting from bracket immersion in the test solution revealed no toxicity on U2OS cells using a methylthiazole tetrazolium (MTT) colorimetric assay. Clearly, the search for an improved technique for enhancing the anticorrosion ability of the normal metal orthodontic bracket should be continued. © 2002 Wiley Periodicals, Inc. J Biomed Mater Res (Appl Biomater) 63: 786–792, 2002
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Institute of Stomatology, Chung Shan Medical and Dental College, 110, Section 1, Chien Kuo N Road, Taichung, Taiwan, PO 402.
Zitieren
APA
Kao, C. T., Ding, S. J., Chen, Y. C., & Huang, T. H. (2002). The anticorrosion ability of titanium nitride (TiN) plating on an orthodontic metal bracket and its biocompatibility. Journal of Biomedical Materials Research, 63(6), 786-792. https://doi.org/10.1002/jbm.10484
Vancouver
Kao CT, Ding SJ, Chen YC, Huang TH. The anticorrosion ability of titanium nitride (TiN) plating on an orthodontic metal bracket and its biocompatibility. Journal of Biomedical Materials Research. 2002;63(6):786-792. doi:10.1002/jbm.10484
BibTeX
@article{Kao2002,
author = {Kao, Chia-Tze and Ding, Shinn-Jyh and Chen, Yu-Chih and Huang, Tsui-Hsien},
title = {The anticorrosion ability of titanium nitride (TiN) plating on an orthodontic metal bracket and its biocompatibility},
journal = {Journal of Biomedical Materials Research},
year = {2002},
volume = {63},
number = {6},
pages = {786--792},
doi = {10.1002/jbm.10484}
}
EUVIMED