Fachartikel in der Zeitschrift Journal of Biophotonics, 2025. Verfasst von Zehua Guan und 7 weiteren Autor:innen. Beteiligte Einrichtung: Shanghai Institute of Technology, College of Sciences, Shanghai, China. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Endoscopy, Angiography. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Science and Technology Innovation Plan Of Shanghai Science and Technology Commission, Collaborative Innovation Project of Shanghai Institute of Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Endoscopic Optical Coherence Tomography (OCT) can provide high‐resolution cross‐sectional images for internal organ tissues. Combining the endoscopic imaging with Optical Coherence Tomography Angiography (OCTA), information of blood vessels in superficial lumen tissues is expected to be acquired. However, in endoscopic OCT systems using proximal scanning probes, performance of OCTA encounters significant challenges due to non‐uniform rotational distortion (NURD) caused by the non‐constant rotation of the distal imaging unit. In this study, we proposed a registration method for endoscopic OCTA imaging in a proximally controlled OCT System. Global registration and A‐line registration were employed to correct the distortion caused by mechanical friction between the catheter sheath and torque coil. Experimental performances in both microfluidic channel and rat rectum show significant correction of NURD. Our study achieved the first implementation of endoscopic OCTA under a spiral B‐scan rotation scheme in a proximally controlled OCT system, facilitating clear visualization of blood flow.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceEndoscopyAngiographyAnimalsRotationRatsImage Processing, Computer-Assisted

Stichwörter der Autor:innen

endoscopic optical coherence tomography angiographynon‐uniform rotational distortion correctionproximal scanning catheter

Autor:innen und Institutionen

  • Shanghai Institute of Technology, College of Sciences, Shanghai, China.
  • Department of Cardiology, Xinhua Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai, China.

ORCID-Profile: Zehua Guan, Shuhao Fan

Förderung

  • National Natural Science Foundation of China
  • Science and Technology Innovation Plan Of Shanghai Science and Technology Commission
  • Collaborative Innovation Project of Shanghai Institute of Technology

Zitieren

APA

Guan, Z., Li, Q., Niu, C., Fan, S., Yu, H., Wu, W., Feng, X., & Dai, C. (2025). Correction of Non‐Uniform Rotational Distortion in the Proximally Controlled Endoscopic OCTA. Journal of Biophotonics. https://doi.org/10.1002/jbio.202400467

Vancouver

Guan Z, Li Q, Niu C, Fan S, Yu H, Wu W, et al. Correction of Non‐Uniform Rotational Distortion in the Proximally Controlled Endoscopic OCTA. Journal of Biophotonics. 2025. doi:10.1002/jbio.202400467

BibTeX
@article{Guan2025,
  author = {Guan, Zehua and Li, Qiang and Niu, Chen and Fan, Shuhao and Yu, Huanhuan and Wu, Wenjuan and Feng, Xiangfei and Dai, Cuixia},
  title = {Correction of Non‐Uniform Rotational Distortion in the Proximally Controlled Endoscopic OCTA},
  journal = {Journal of Biophotonics},
  year = {2025},
  doi = {10.1002/jbio.202400467}
}
Kostenlos starten →