Multizentrische Studie in der Zeitschrift International Journal of Gynecology & Obstetrics, 2025. Verfasst von Michal Yaron und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Woman, Child & Adolescent, University Hospitals (HUG), Genève, Switzerland. Hauptthemen laut MEDLINE: Intrauterine Devices. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Aspivix SA.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To evaluate the experiences with an atraumatic device (Carevix®) for intrauterine device (IUD) placement in routine clinical practice.

Methods

Our prospective, post‐marketing study was conducted in 19 centers in Europe, Brazil and the USA. IUD insertions using Carevix device were documented for completion rates, patient‐reported pain scores, cervical bleeding, and operator and patient satisfaction. Demographics, uterine position, and menstrual cycle phase were also recorded.

Results

A total of 1123 IUD placements were analyzed. Most participants (819/992; 83%) were 18–40 years old; 46% (513/1123) were nulliparous and 47% (526/1123) were menstruating at the time of the procedure. The procedure was completed with Carevix alone in 82% (925/1123) of participants. Additional tenaculum use was mostly due to spontaneous vacuum loss (194/1123). Completion rates without a tenaculum were lower in nulliparous than in parous women (78% versus 86%). Average visual analog pain scores were lower with Carevix alone (35 ± 21) than when a tenaculum was required (43 ± 21). Most participants (73%) reported less pain than expected, and only 3% reported more pain. Among providers, 82% were satisfied with Carevix, citing ease of use, visibility, and low patient discomfort.

Conclusions

In a multicenter real‐world setting, the novel, atraumatic device demonstrated feasibility and safety, with high completion rates and high satisfaction among both operators and patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intrauterine DevicesHumansFemaleAdultProspective StudiesYoung AdultProduct Surveillance, PostmarketingPatient SatisfactionAdolescentBrazilEuropeUnited StatesCervix UteriPain Measurement

Stichwörter der Autor:innen

bleedingintrauterine devicenulliparouspainreal‐life studysatisfactionsuction cervical stabilizertranscervical

Autor:innen und Institutionen

  • Department of Woman, Child & Adolescent, University Hospitals (HUG), Genève, Switzerland.
  • Department of Obstetrics and Gynecology, University of Campinas Faculty of Medical Sciences, Campinas, SP, Brazil.
  • University Obstetricians-Gynecologists Coleman Center, Indianapolis, Indiana, USA.

Förderung

  • Aspivix SA

Zitieren

APA

Yaron, M., Crofts, V., Marcelino, A. C., Espejo-Arce, X., Garcia, E., Barbe, J., Vilaplana, E. G., Murtas, M., Juliato, C. T., Moussaoui, D., Conklin, A., & Bahamondes, L. (2025). Safety and efficacy of a suction cervical stabilizer for IUD insertion: Results from a multicenter post‐marketing study. International Journal of Gynecology & Obstetrics. https://doi.org/10.1002/ijgo.70718

Vancouver

Yaron M, Crofts V, Marcelino AC, Espejo-Arce X, Garcia E, Barbe J, et al. Safety and efficacy of a suction cervical stabilizer for IUD insertion: Results from a multicenter post‐marketing study. International Journal of Gynecology & Obstetrics. 2025. doi:10.1002/ijgo.70718

BibTeX
@article{Yaron2025,
  author = {Yaron, Michal and Crofts, Victoria and Marcelino, Ana C and Espejo-Arce, Ximena and Garcia, Elaine and Barbe, Julia and Vilaplana, Elisabeth Garcia and Murtas, Micol and Juliato, Cassia T and Moussaoui, Dehlia and Conklin, Alissa and Bahamondes, Luis},
  title = {Safety and efficacy of a suction cervical stabilizer for IUD insertion: Results from a multicenter post‐marketing study},
  journal = {International Journal of Gynecology & Obstetrics},
  year = {2025},
  doi = {10.1002/ijgo.70718}
}
Kostenlos starten →