The impact of metal cup size on neonatal and maternal morbidity in vacuum‐assisted deliveries
International Journal of Gynecology & Obstetrics · 2025
DOI: 10.1002/ijgo.70667 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Gynecology & Obstetrics, 2025. Verfasst von Matan Anteby und 6 weiteren Autor:innen. Beteiligte Einrichtung: Lis Hospital for Women's Health, Tel Aviv Sorasky Medical Center, Gray Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel. Hauptthemen laut MEDLINE: Vacuum Extraction, Obstetrical, Birth Injuries. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
To evaluate whether metal cup size influences maternal and neonatal morbidity in vacuum‐assisted deliveries (VAD). Materials and Methods Retrospective cohort at a single tertiary center (2011–2022) comparing 60‐mm versus 50‐mm Malmström cups. A 1:1 matched cohort analysis balanced maternal and obstetric variables. Primary neonatal outcomes were a birth trauma composite outcome (brachial plexus injury, cephalohematoma, subgaleal or subdural hematoma) and a severe neonatal composite outcome (neonatal intensive care unit admission, 5‐min Apgar score less than 7, umbilical artery pH less than 7.1, intraventricular hemorrhage, hypoxic ischemic encephalopathy, or neonatal convulsions). The primary maternal composite outcome included postpartum hemorrhage, third‐ or fourth‐degree perineal tears or need for postpartum blood products.
Results
After matching, 1564 patients were included (782 per group). The 60‐mm cup was associated with fewer cup detachments (22 [2.8%] versus 41 [5.2%], P = 0.015), a shorter interval to delivery (4.3 ± 1.9 versus 4.7 ± 2.4 min, P = 0.003) and a comparable failure rate leading to cesarean delivery (5 [0.6%] versus 7 [0.9%], P = 0.56). The neonatal birth trauma composite outcome was less prevalent with 60‐mm cups (51 [6.5%] versus 72 [9.2%]; odds ratio [OR] 0.69, 95% confidence interval [CI] 0.47–0.99, P = 0.049) whereas the severe neonatal composite outcome did not differ (83 [10.6%] versus 96 [12.3%], P = 0.302). Maternal composite outcome was similar between groups; however, perineal tears were less common with 60‐mm cups (306 [39.1%] vs 369 [47.2%]; OR 0.72, 95% CI 0.59–0.88; P = 0.001).
Conclusion
In VAD, use of a larger Malmström cup (60 mm) is associated with fewer cup detachments and a reduced risk of neonatal birth trauma without increasing maternal morbidity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Lis Hospital for Women's Health, Tel Aviv Sorasky Medical Center, Gray Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Zitieren
APA
Anteby, M., Pinchas-Cohen, T., Baruch, Y., Lavie, A., Maslovitz, S., Hiersch, L., & Yogev, Y. (2025). The impact of metal cup size on neonatal and maternal morbidity in vacuum‐assisted deliveries. International Journal of Gynecology & Obstetrics. https://doi.org/10.1002/ijgo.70667
Vancouver
Anteby M, Pinchas-Cohen T, Baruch Y, Lavie A, Maslovitz S, Hiersch L, et al. The impact of metal cup size on neonatal and maternal morbidity in vacuum‐assisted deliveries. International Journal of Gynecology & Obstetrics. 2025. doi:10.1002/ijgo.70667
BibTeX