EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

In Silico Modeling of Safe Force Limits for Trismus Therapy Following Jaw Reconstruction

Masoud Mohseni-Dargah iD, Boyang Wan, Emma Charters iD, Masako Dunn, Timothy Manzie, Yee Mon Aung, Shaheen Hasmat, Qing Li, Jonathan R Clark

Head & Neck · 2025-11 · 47 (11) · 3115-3122

DOI: 10.1002/hed.28230 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Head & Neck, online erschienen am 02.07.2025. Verfasst von Masoud Mohseni-Dargah und 8 weiteren Autor:innen. Beteiligte Einrichtung: AMME School, Engineering Faculty, The University of Sydney, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Trismus, Mandibular Reconstruction. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: The Cancer Institute New South Wales, Sydney Local Health District, The Walker Family Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Objective Restricted mouth opening (trismus) is a common complication following segmental mandibulectomy and autologous bone reconstruction. However, the biomechanical limits of reconstructed mandibles, particularly in cases without bone union, remain poorly understood. Restorabite is a novel jaw‐stretching device designed to increase the safety of trismus therapy by controlling the applied force. This study estimates the patient‐specific loading limits in compromised mandibular reconstruction using finite element analysis to guide safe trismus therapy.

Methods

Based on patient imaging, the finite element method was used to model stress distributions in the bone, titanium plate, and screws of a reconstructed mandible lacking bone union. Various scenarios were simulated to determine safe loading parameters, which considered factors such as friction coefficients between bone segments, native bone, and the reconstructed plate, as well as force application sites, and concentrated versus distributed loading.

Results

Low friction coefficients significantly increased stress on the reconstruction plate. To simulate the worst‐case scenario, a coefficient of 0.3 was applied in the subsequent simulations. Stress levels varied by loading sites, with concentrated loading on the posterior segment of grafted bone yielding the lowest stress. In contrast, distributed loading over the 4 cm region on the anterior segments of graft bone led to increased plate stress.

Conclusions

This study highlights the importance of tailored approaches in trismus therapy to minimize osteolysis and plate failure in mandibular reconstruction providing valuable insights for future clinical practices and device development. The study found that concentrated loading on biomechanically favorable regions proved safer compared to distributed loading, as it occurs with traditional jaw‐stretching devices.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TrismusMandibular ReconstructionHumansFinite Element AnalysisComputer SimulationBone PlatesBiomechanical PhenomenaBone ScrewsStress, MechanicalMandible

Stichwörter der Autor:innen

bone remodelingfinite element analysisfrictionmandibular osteotomytrismus

Autor:innen und Institutionen

  • AMME School, Engineering Faculty, The University of Sydney, Sydney, New South Wales, Australia.
  • Department of Head and Neck Surgery, Chris O'brien Lifehouse, Sydney, New South Wales, Australia.
  • Sydney Medical School, Faculty of Medicine and Health, University of Sydney, Sydney, New South Wales, Australia.
  • Central Clinical School, Faculty of Medicine and Health Sciences, The University of Sydney, Sydney, New South Wales, Australia.
  • Sydney School of Health Sciences, The University of Sydney, Sydney, New South Wales, Australia.
  • NHMRC Centre of Research Excellence for Applied Innovations in Oral Cancer, The University of Sydney, Sydney, Australia.

ORCID-Profile: Masoud Mohseni-Dargah, Emma Charters

Förderung

  • The Cancer Institute New South Wales
  • Sydney Local Health District
  • The Walker Family Foundation
  • Surfebruary
  • Australian and New Zealand Head and Neck Cancer Society
  • the Royal Australasian College of Surgeons
  • the National Health and Medical Research Council
  • The Australian and New Zealand Association of Oral and Maxillofacial Surgeons (ANZAOMS)

Zitieren

APA

Mohseni-Dargah, M., Wan, B., Charters, E., Dunn, M., Manzie, T., Aung, Y. M., Hasmat, S., Li, Q., & Clark, J. R. (2025). In Silico Modeling of Safe Force Limits for Trismus Therapy Following Jaw Reconstruction. Head & Neck, 47(11), 3115-3122. https://doi.org/10.1002/hed.28230

Vancouver

Mohseni-Dargah M, Wan B, Charters E, Dunn M, Manzie T, Aung YM, et al. In Silico Modeling of Safe Force Limits for Trismus Therapy Following Jaw Reconstruction. Head & Neck. 2025;47(11):3115-3122. doi:10.1002/hed.28230

BibTeX
@article{MohseniDargah2025,
  author = {Mohseni-Dargah, Masoud and Wan, Boyang and Charters, Emma and Dunn, Masako and Manzie, Timothy and Aung, Yee Mon and Hasmat, Shaheen and Li, Qing and Clark, Jonathan R},
  title = {In Silico Modeling of Safe Force Limits for Trismus Therapy Following Jaw Reconstruction},
  journal = {Head & Neck},
  year = {2025},
  volume = {47},
  number = {11},
  pages = {3115--3122},
  doi = {10.1002/hed.28230}
}
Kostenlos starten →