EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

A Large‐Area, Stretchable Accelerometer Array for Spatial Assessment of Chest Compressions in Cardiopulmonary Resuscitation

Tianrun Xu, Jin Ji, Chongyuan Huang, Yiming Chen, Peng Zhao, Jizhou Song

FlexTech · 2026

DOI: 10.1002/fle2.70009 ↗

Kurz gefasst

Fachartikel in der Zeitschrift FlexTech, 2026. Verfasst von Tianrun Xu und 5 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Key Research and Development Program of China, National Natural Science Foundation of China.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

High‐quality chest compressions are vital for successful cardiopulmonary resuscitation (CPR). Existing manual and mechanical compression methods suffer from unstable quality, lack of real‐time feedback, and insufficient personalized adjustment. Here, we report a large‐area, stretchable accelerometer array with the capability to monitor multi‐point compression depth across the chest and achieve the cross‐sectional chest contour at the peak compression. The proposed accelerometer array features stretchable discrete stiff island design and exhibits good mechanical adaptability to enable conformal contact with the chest region. A chest compression depth estimation algorithm based on triaxial acceleration data is developed to predict the compression depth. Moreover, a peak‐averaging‐based contour construction method is proposed to construct the deformed chest contour. The predicted compression depth and chest contour agree well with experimental measurements in both simple hollow semi‐cylinder and complex adult manikin systems, demonstrating that the accelerometer array exhibits high spatiotemporal resolution in multi‐point chest compression monitoring and dynamic tracking of thoracic deformation patterns. These results offer high spatiotemporal resolution for multi‐point deformation tracking, providing key technical support for advancing the CPR feedback technologies.

Förderung

  • National Key Research and Development Program of China
  • National Natural Science Foundation of China

Zitieren

APA

Xu, T., Ji, J., Huang, C., Chen, Y., Zhao, P., & Song, J. (2026). A Large‐Area, Stretchable Accelerometer Array for Spatial Assessment of Chest Compressions in Cardiopulmonary Resuscitation. FlexTech. https://doi.org/10.1002/fle2.70009

Vancouver

Xu T, Ji J, Huang C, Chen Y, Zhao P, Song J. A Large‐Area, Stretchable Accelerometer Array for Spatial Assessment of Chest Compressions in Cardiopulmonary Resuscitation. FlexTech. 2026. doi:10.1002/fle2.70009

BibTeX
@article{Xu2026,
  author = {Xu, Tianrun and Ji, Jin and Huang, Chongyuan and Chen, Yiming and Zhao, Peng and Song, Jizhou},
  title = {A Large‐Area, Stretchable Accelerometer Array for Spatial Assessment of Chest Compressions in Cardiopulmonary Resuscitation},
  journal = {FlexTech},
  year = {2026},
  doi = {10.1002/fle2.70009}
}
Kostenlos starten →