Artificial intelligence ‐assisted detection of epileptic spasms using electroencephalographic–video analysis
Epilepsia · 2026-06 · 67 (6) · 3009-3022
DOI: 10.1002/epi.70194 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Epilepsia, online erschienen am 31.03.2026. Verfasst von Lin Wan und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, First Medical Center, Chinese PLA (People's Liberation Army) General Hospital, Beijing, China. Hauptthemen laut MEDLINE: Artificial Intelligence, Electroencephalography, Video Recording, Epilepsy. Die Arbeit nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: general project of the National Key Research and Development Program of China, key project of innovation cultivation fund of the Seventh Medical Center of Chinese PLA General Hospital, Capital's Funds for Health Improvement and Research.
Abstract (Originalsprache)
Objective
This study was undertaken to develop and validate an artificial intelligence (AI) diagnostic tool using hybrid electroencephalographic (EEG)–video signals for automatic epileptic spasms (ES) detection.
Methods
This retrospective cohort study with internal cross‐validation and multicenter external validation was conducted from July 2022 to May 2025. It included 252 patients with ES from Chinese PLA General Hospital and 60 from three other medical centers. We developed a multimodal fusion approach combining video and electrophysiological signals. All EEG data were segmented into continuous 4‐s pages. The internal cohort consisted of 212 patients (723.4 h video‐EEG, 7348 ES, and 643 215 non‐ES segments). Clinical validation involved 100 patients across four datasets (212.2 h, 5709 ES and 185 207 non‐ES segments) plus 78 controls without ES (218.1 h, 196 249 segments, used for false alarm rate analysis). Primary outcomes included sensitivity, specificity, accuracy, precision, and F1 score versus electroencephalographer interpretation.
Results
In internal cross‐validation, the hybrid EEG–video model achieved superior performance compared to current EEG‐only models (area under the precision–recall curve = .7334, 95% confidence interval [CI] = .7118–.7539, area under the receiver operating characteristic curve = .9820, 95% CI = .9782–.9856). In the clinical validation dataset, the model demonstrated diagnostic sensitivity (.735, 95% CI = .673–.822) and specificity (.995, 95% CI = .992–.996) comparable to experienced electroencephalographers. With AI assistance, all electroencephalographers showed improved sensitivity trends, with one rater showing a marked improvement from .620 (95% CI = .532–.699) to .792 (95% CI = .757–.821). Notably, specificity did not decline significantly for any of the raters. In samples containing subtle ES, machine‐assisted recognition significantly improved sensitivity by 16%–21% for all electroencephalographers ( p < .01). The model maintained low false alarm rates (.16‰) across different patient populations including healthy controls and epilepsy patients without ES. Significance This AI diagnostic tool achieves clinical performance comparable to senior electroencephalographers when applied independently for ES detection, with greater robustness in detecting subtle ES. When used collaboratively with clinicians, it could enhance diagnostic sensitivity while maintaining high specificity. The technology addresses critical diagnostic challenges in pediatric epilepsy care and shows promise to reduce health care inequities in resource‐limited settings lacking specialized expertise.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pediatrics, First Medical Center, Chinese PLA (People's Liberation Army) General Hospital, Beijing, China.
- Senior Department of Pediatrics, Seventh Medical Center of Chinese PLA General Hospital, Beijing, China.
- Department of Neurology, Peking Union Medical College Hospital, Beijing, China.
- NetEase Media Technology Co., Beijing, China.
- Department of Pediatrics, Shengjing Hospital of China Medical University, Shenyang, China.
- Department of Pediatrics, Peking University First Hospital, Beijing, China.
ORCID-Profile: Lin Wan, Nan Lin, Wen Wang, Zi Liang, Yisu Dong, Jian Chen, Yuying Fan, Tong Liu, Hongjie Wang, Weike Cheng, Guangshuang Lu, Juan Wang, Bo Zhang, Qiang Lu, Guang Yang
Förderung
- general project of the National Key Research and Development Program of China
- key project of innovation cultivation fund of the Seventh Medical Center of Chinese PLA General Hospital
- Capital's Funds for Health Improvement and Research
- Epilepsy research foundation of Chinese Association Against Epilepsy
- Innovation Talent Fund of Senior Department of Pediatrics, The Seventh Medical Center of PLA General Hospital
Zitieren
APA
Wan, L., Lin, N., Wang, W., Liang, Z., Dong, Y., He, H., Chen, J., Fan, Y., Liu, T., Wang, H., Cheng, W., Lu, G., Wang, J., Zhang, B., Lu, Q., & Yang, G. (2026). Artificial intelligence ‐assisted detection of epileptic spasms using electroencephalographic–video analysis. Epilepsia, 67(6), 3009-3022. https://doi.org/10.1002/epi.70194
Vancouver
Wan L, Lin N, Wang W, Liang Z, Dong Y, He H, et al. Artificial intelligence ‐assisted detection of epileptic spasms using electroencephalographic–video analysis. Epilepsia. 2026;67(6):3009-3022. doi:10.1002/epi.70194
BibTeX
@article{Wan2026,
author = {Wan, Lin and Lin, Nan and Wang, Wen and Liang, Zi and Dong, Yisu and He, Haibo and Chen, Jian and Fan, Yuying and Liu, Tong and Wang, Hongjie and Cheng, Weike and Lu, Guangshuang and Wang, Juan and Zhang, Bo and Lu, Qiang and Yang, Guang},
title = {Artificial intelligence ‐assisted detection of epileptic spasms using electroencephalographic–video analysis},
journal = {Epilepsia},
year = {2026},
volume = {67},
number = {6},
pages = {3009--3022},
doi = {10.1002/epi.70194}
}
EUVIMED