EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment

Kyle D Luttgeharm iD, Manu Grover iD, Sheng-Yuan Huang iD, Whitney A Pike iD

ELECTROPHORESIS · 30.10.2026

DOI: 10.1002/elps.70174 ↗

Kurz gefasst

Fachartikel in der Zeitschrift ELECTROPHORESIS, online erschienen am 30.10.2026. Verfasst von Kyle D Luttgeharm und 3 weiteren Autor:innen. Beteiligte Einrichtung: Agilent Technologies, Ankeny, Iowa, USA. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Fluorescent capillary gel electrophoresis (CGE) with sodium dodecyl sulfate (CE‐SDS) provides a powerful, high‐sensitivity alternative to ultraviolet (UV)‐based detection for characterizing therapeutic monoclonal antibodies (mAbs). Regulatory and standards organizations, such as the United States Pharmacopeia (USP), include only UV‐based CE‐SDS methods, hindering the adoption of alternative detection methods. There is growing opportunity to expand beyond exclusively UV‐based CE‐SDS methods. In this study, we present an analytical validation of a light‐emitting diode (LED) fluorescence‐based parallel CE‐SDS method for both non‐reduced and reduced analysis of therapeutic antibodies. Using the NISTmAb reference material as a model system, size heterogeneity critical quality attributes (CQAs), including monomeric purity (non‐reduced conditions), percent glycosylation (reduced conditions), and percent thioether (reduced conditions), were assessed. The fluorescence method demonstrated high specificity and precision with relative standard deviation (RSD) values

Schlagwörter

Stichwörter der Autor:innen

antibody critical quality attributesantibody method validationcapillary electrophoresis SDSmonoclonal antibody

Autor:innen und Institutionen

  • Agilent Technologies, Ankeny, Iowa, USA.
  • Agilent Technologies, Santa Clara, California, USA.

ORCID-Profile: Kyle D Luttgeharm, Manu Grover, Sheng-Yuan Huang, Whitney A Pike

Zitieren

APA

Luttgeharm, K. D., Grover, M., Huang, S. Y., & Pike, W. A. (2026). Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment. ELECTROPHORESIS. https://doi.org/10.1002/elps.70174

Vancouver

Luttgeharm KD, Grover M, Huang SY, Pike WA. Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment. ELECTROPHORESIS. 2026. doi:10.1002/elps.70174

BibTeX
@article{Luttgeharm2026,
  author = {Luttgeharm, Kyle D and Grover, Manu and Huang, Sheng-Yuan and Pike, Whitney A},
  title = {Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment},
  journal = {ELECTROPHORESIS},
  year = {2026},
  doi = {10.1002/elps.70174}
}
Kostenlos starten →