Improved separation of fentanyl isomers using metal cation adducts and high‐resolution ion mobility‐mass spectrometry

Verbesserte Trennung von Fentanyl-Isomeren mit Metallkation-Addukten und hochauflösender Ionenmobilitäts-Massenspektrometrie

Ralph Aderorho, Christopher D Chouinard iD

Drug Testing and Analysis · 2024-04 · 16 (4) · 369-379

DOI: 10.1002/dta.3550 ↗

Auf einen Blick

Die Studie entwickelte eine schnelle IM-MS-Methode zur Detektion und Trennung von Fentanyl-Isomeren. [M + Cu]+- und [M + Ag]+-Addukte verbesserten die Auflösung deutlich.

Fragestellung

Lässt sich Fentanyl mit IM-MS und Metallkation-Addukten schneller und besser zwischen Isomeren unterscheiden?

Methode

Entwicklung einer IM-MS-Methode; Einsatz ungewöhnlicher Kation-Addukte und hochauflösender Demultiplexierung; Sammlung von CCS-Werten für verschiedene Metallionen-Addukte.

Ergebnisse

[M + Cu]+- und [M + Ag]+-Addukte verbesserten die Auflösung zwischen Isomeren. Mit hochauflösender Nachbearbeitung wurde eine Trennleistung von >150 erreicht; die Methode quantifizierte Fentanyl im Humanurin mit LOD 13 pg/mL und LOQ 40 pg/mL.

Schlussfolgerung der Autor:innen

Die verbesserten Trennstrategien erlaubten eine sichere Detektion und anschließende quantitative Analyse.

Kernaussagen

  • IM-MS zur Trennung von Fentanyl-Isomeren
  • [M + Cu]+ und [M + Ag]+ verbesserten die Auflösung
  • Humanurin: LOD 13 pg/mL, LOQ 40 pg/mL

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Drug Testing and Analysis, online erschienen am 25.07.2023. Verfasst von Ralph Aderorho und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemistry, Clemson University, Clemson, SC, USA. Hauptthemen laut MEDLINE: Fentanyl, Ion Mobility Spectrometry. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 69 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Clemson University.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Fentanyl is a potent synthetic opioid that has attracted significant attention due to its illegal production and distribution, resulting in misuse, overdose, and fatalities. Because numerous fentanyl analogs, including structural isomers, with different potency have been discovered in the field, there is a critical need to continue developing analytical methodologies capable of accurate identification in forensic and clinical laboratories. This study aimed to develop a rapid method for detecting and separating fentanyl isomers based on ion mobility–mass spectrometry (IM‐MS), where IM separates gas‐phase ions based on differences in their size, shape, and charge. Several strategies for improved differentiation were implemented, including using unconventional cation adducts (e.g., alkali and transition metals) and data post‐processing by high‐resolution demultiplexing. A collection of collision cross section (CCS) values for the various metal ion adducts was gathered, which can be used to improve confidence of identification in future samples. Notable examples, such as [M + Cu] + and [M + Ag] + adducts, contributed to significant improvement of resolution between isomers. Furthermore, the addition of high‐resolution post‐processing provided resolving power of >150, which constitutes a significant increase in comparison with the normal 50–60 obtained with low‐resolution drift tube instruments. Collectively, these improved separation strategies allowed for confident detection and subsequent quantitative analysis. The optimized IM‐MS method resulted in quantification of fentanyl in human urine with limits of detection and quantification of 13 pg/mL and 40 pg/mL, respectively.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

FentanylIon Mobility SpectrometryHumansCationsIsomerismMass Spectrometry

Stichwörter der Autor:innen

fentanylion mobility–mass spectrometry (IM‐MS)isomers

Substanzen

FentanylCations

Autor:innen und Institutionen

  • Department of Chemistry, Clemson University, Clemson, SC, USA.

ORCID-Profile: Christopher D Chouinard

Förderung

  • Clemson University

Zitieren

APA

Aderorho, R., & Chouinard, C. D. (2024). Improved separation of fentanyl isomers using metal cation adducts and high‐resolution ion mobility‐mass spectrometry. Drug Testing and Analysis, 16(4), 369-379. https://doi.org/10.1002/dta.3550

Vancouver

Aderorho R, Chouinard CD. Improved separation of fentanyl isomers using metal cation adducts and high‐resolution ion mobility‐mass spectrometry. Drug Testing and Analysis. 2024;16(4):369-379. doi:10.1002/dta.3550

BibTeX
@article{Aderorho2024,
  author = {Aderorho, Ralph and Chouinard, Christopher D},
  title = {Improved separation of fentanyl isomers using metal cation adducts and high‐resolution ion mobility‐mass spectrometry},
  journal = {Drug Testing and Analysis},
  year = {2024},
  volume = {16},
  number = {4},
  pages = {369--379},
  doi = {10.1002/dta.3550}
}
Kostenlos starten →