Preclinical Evaluation of the Safety of the Dedicated Gel for Gel Immersion Endoscopy With Gastrointestinal Perforation and Peritoneal Leakage
DEN Open · 2026-04 · 6 (1)
DOI: 10.1002/deo2.70258 ↗Kurz gefasst
Fachartikel in der Zeitschrift DEN Open, online erschienen am 08.12.2025. Verfasst von Atsushi Ohata und 5 weiteren Autor:innen. Beteiligte Einrichtung: Naruto Research Institute Research and Development Center Otsuka Pharmaceutical Factory Inc. Tokushima Japan. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Gel immersion endoscopic mucosal resection or endoscopic submucosal dissection has been developed as an alternative to saline immersion therapeutic endoscopy. However, in cases of gastrointestinal perforation, gel may leak from the lumen into the peritoneal cavity, and the safety of such events remains unclarified. This study aimed to evaluate the safety of the dedicated gel (Viscoclear, VC) in the event of perforation and peritoneal leakage.
Methods
A perforation model was created to compare the colonic burst pressure after VC or saline injection. Acute toxicity was assessed by intraperitoneal VC administration in rats at a maximum dose of 20 mL/kg. Escherichia coli was cultured for 24 h in VC or saline, and bacterial growth was evaluated. Cytotoxicity (colony formation of V79 cells), sensitization (guinea pig maximization test), and intracutaneous reactivity (local irritation in rabbits) were also examined.
Results
The burst pressure was significantly higher for VC injection than for saline. There were no abnormalities indicative of toxicity observed in the general condition or at necropsy after VC administration, and the hematological and biochemical parameters were within normal limits. VC did not promote bacterial growth compared with saline. VC was not cytotoxic and did not cause skin sensitization. Intradermal reactivity tests showed that VC caused negligible irritation that was judged “acceptable”.
Conclusions
These results suggest that VC is less likely to leak than saline and that, even in the event of a minor leakage into the peritoneal cavity, VC has no acute systemic toxicity and acceptable biocompatibility.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Naruto Research Institute Research and Development Center Otsuka Pharmaceutical Factory Inc. Tokushima Japan.
- Department of Medicine Division of Gastroenterology Jichi Medical University Tochigi Japan.
ORCID-Profile: Tomonori Yano
Zitieren
APA
Ohata, A., Kawahara, R., Hiraki, Y., Nakata, H., Kaneda, S., & Yano, T. (2026). Preclinical Evaluation of the Safety of the Dedicated Gel for Gel Immersion Endoscopy With Gastrointestinal Perforation and Peritoneal Leakage. DEN Open, 6(1). https://doi.org/10.1002/deo2.70258
Vancouver
Ohata A, Kawahara R, Hiraki Y, Nakata H, Kaneda S, Yano T. Preclinical Evaluation of the Safety of the Dedicated Gel for Gel Immersion Endoscopy With Gastrointestinal Perforation and Peritoneal Leakage. DEN Open. 2026;6(1). doi:10.1002/deo2.70258
BibTeX
@article{Ohata2026,
author = {Ohata, Atsushi and Kawahara, Ryo and Hiraki, Yuji and Nakata, Hikaru and Kaneda, Shinya and Yano, Tomonori},
title = {Preclinical Evaluation of the Safety of the Dedicated Gel for Gel Immersion Endoscopy With Gastrointestinal Perforation and Peritoneal Leakage},
journal = {DEN Open},
year = {2026},
volume = {6},
number = {1},
doi = {10.1002/deo2.70258}
}
EUVIMED